Operator responsibility – Safety instructions
8
2.2. Technical specifications
HT30-M1
HT30-M1-C.U.
Working temperature range
°C
70 ... 400
40 … 400
Temperature stability external
°C
±0,01 ... ±0.1
Display accuracy
%
±0.5 ±1Digit
Temperature selection
digital
via keypad
indication on LCD DIALOG-DISPLAY (°C/°F)
remote control via PC
indication on monitor
Temperature indication
Multi-Display (LED)
DIALOG-Display (LCD)
Resolution
°C
0.01
Absolute Temperature Calibration
(ATC 1)
°C
±9.99
(ATC 2)
°C
±9.99
Temperature control
ICC - Intelligent Cascade Control,
self-optimizing
Cascade, parameter can be called-in and modified
Heater wattage (at 230 V)
kW
3.1
C.U.-cooling unit:
Temperature
<IntAct>
Cooling capacity
at 2 l/min flow of cooling water
°C
kW
-----
350 250
150
75
-----
12
8
4
1
Pressure pump:
Max. pressure
max. at 0 liters
bar
0.8 … 1.2
Flow rate
max. at 0 bar
l/min
14 … 18
Electrical connections:
See page 9
Filling volume
litre
2
Overall dimensions (WxDxH)
M1
Cm
25x25x18
25x25x18
Overall dimensions (WxDxH)
HT30
Cm
23x23x58
43x23x58
Weight
kg
27
35
Ambient temperature
°C
5 ... 40
Mains power connection 230 V/50 HZ
V / Hz
230 10 % / 50
Current draw
(at 230 V)
A
15
Mains power connection 230 V/60 HZ
V / Hz
230 10 % / 60
Current draw
(at 230 V)
A
15
All measurements have been carried out at rated voltage and frequency, ambient temperature: 20 °C
.