31
7.6.
AUTOSTART ON / OFF
The recirculating cooler has been configured and supplied by JULABO
according to N.A.M.U.R. recommendations. This means for the start mode,
that the unit must enter a safe operating state after a power failure (non-
automatic start mode). This safe operating state is indicated by „OFF“ on the
LED temperature display. A complete shutdown of the main functional
elements such as compressor and circulating pump is effected
simultaneously.
Should such a safety standard not be required, the AUTOSTART function
(automatic start mode) may be activated, thus allowing the start of the
circulator directly by pressing the mains power switch or using a timer.
Keep depressed enter
and
turn on the unit with the mains power switch.
For a short while the LED DISPLAY indicates the effective start mode:
AUTOSTART on.
AUTOSTART off.
Warning:
For supervised or unsupervised operation with the AUTOSTART function, avoid any
hazardous situation to persons or property.
The circulator does no longer conform to N.A.M.U.R. recommendations.