33
8. Safety
installations
8.1. Excess
temperature protection
+
This safety installation is independent of the control circuit.
When the temperature of the bath fluid has reached the safety temperature
(85 °C), a complete shutdown of the compressor and pump is effected.
The alarm is indicated by optical and audible signals (continuous tone) and on
the LED-DISPLAY appears the error message
"
E
rror
14
".
8.2.
Low level protection
+
This safety installation is independent of the control circuit.
If the low liquid level protection device is triggered, a complete shutdown of
the compressor and circulating pump is effected.
The alarm is indicated by optical and audible signals (continuous tone) and on
the LED-DISPLAY appears the error message
"
E
rror
01
".
Turn off the unit with the mains switch, refill bath fluid and turn the unit on
again!
Caution:
For refill always use the same bath fluid type that is already in the bath.