35
The return temperature is above the switch-off value of the high temperature
protection (85°C). Check dimensioning of application.
Use a stronger recirculating cooler if necessary.
After eliminating the malfunction, press the mains power switch off and on
again to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized service
station.
Warning without a complete shutdown of the unit
Excess temperature warning starting at 75 °C
The return temperature soon reaches the swith-off value of the high
temperature protection (85 °C).
If the unit cannot be returned to operation, contact an authorized JULABO
service station.
10
A
M
P
Disturbances that are not indicated.
Overload protection:: a) for cooling machine
b) for pump motor
After a short cooling interval, the unit will automatically start running.
Mains circuit breakers (resettable) –10A.