Remote control
92
Command
Parameter
Response of circulator
IN_MODE_01
none
Selected setpoint:
0 = Setpoint 1
1 = Setpoint 2
2 = Setpoint 3
IN_MODE_02
none
Selftuning type:
0 = Selftuning “off”
1 = Selftuning “once”
2 = Selftuning “always”
IN_MODE_03
none
Type of external programmer input:
0 = Voltage 0 V to 10 V
1 = Current 0 mA to 20 mA
IN_MODE_04
none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
IN_MODE_05
none
Circulator in stop/start condition:
0 = stop
1 = start
IN_MODE_08
none
Adjusted control dynamics
0 = aperiodic
1 = standard
IN_HIL_00
none
Max. cooling power (%).
IN_HIL_01
none
Max. heating power (%).
12.4. Status messages
Status messages
Description
00 MANUAL STOP
Circulator in „OFF“ state.
01 MANUAL START
Circulator in keypad control mode.
02 REMOTE STOP
Circulator in „r OFF“ state.
03 REMOTE START
Circulator in remote control mode.
12.5. Error messages
Error messages
Description
-01 LOW LEVEL ALARM
Low liquid level alarm
-02 REFRIGERATOR ALARM
Control cable of the refrigerated circulator or MVS
solenoid valve controller short-circuited or disconnected.
-03 EXCESS TEMPERATURE
WARNING
High temperature warning
-04 LOW TEMPERATURE WARNING
Low temperature warning.