![Julabo F81-HL Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/julabo/f81-hl/f81-hl_operating-manual_2030729081.webp)
HL, SL
81
9.9.5.
Alarm-output
ALARM
Alarm output (10)
(for external alarm signal)
This socket is a potential-free change-over contact. With the
adjustments in the menu >Inputs/Outputs<, all operating conditions
can be signaled without having to change the pin assignments.
Meaning of the terms under menu item >Function< :
The circulator is in condition
>StandBy<
or
>Alarm<
For >Type: normal<, pins 2 and 3 are connected in any case
according to the selected >Function<.
For >Type: inverse<, pins 2 and 1 are connected in any case
according to the selected >Function<.
Switching capacity max. 30 W/40 VA
Switching voltage max. 125 V
∼
/
−
Switching current
max. 1 A