![Julabo F81-HL Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/julabo/f81-hl/f81-hl_operating-manual_2030729091.webp)
HL, SL
91
Command
Parameter
Response of circulator
IN_PV_02
none
Temperature value registered by the external Pt100 sensor.
IN_PV_03
none
Temperature value registered by the safety sensor.
IN_PV_04
none
Setpoint temperature (“SafeTemp”) of the excess
temperature protection
IN_SP_00
none
Working temperature “Setpoint 1”
IN_SP_01
none
Working temperature “Setpoint 2”
IN_SP_02
none
Working temperature “Setpoint 3”
IN_SP_03
none
High temperature warning limit
“OverTemp”
IN_SP_04
none
Low temperature warning limit
“SubTemp”
IN_SP_05
none
Setpoint temperature of the external programmer
(REG+E-PROG).
IN_SP_06
none
Temperature indication in °C or °F
IN_SP_07
none
1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotational
speed after start.
IN_SP_08
none
Value of a flow rate sensor connected to the E-Prog input
IN_SP_09
none
Value of a pressure sensor connected to the E-Prog input
IN_PAR_00
none
Temperature difference between working sensor and safety
sensor
IN_PAR_01
none
Te - Time constant of the external bath.
IN_PAR_02
none
Si - Internal slope
IN_PAR_03
none
Ti - Time constant of the internal bath.
IN_PAR_04
none
CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
IN_PAR_05
none
Factor pk/ph0:
Ratio of max. cooling capacity to max. heating capacity
IN_PAR_06
none
Xp control parameter of the internal controller.
IN_PAR_07
none
Tn control parameter of the internal controller.
IN_PAR_08
none
Tv control parameter of the internal controller.
IN_PAR_09
none
Xp control parameter of the cascade controller.
IN_PAR_10
none
Proportional share of the cascade controller.
IN_PAR_11
none
Tn control parameter of the cascade controller.
IN_PAR_12
none
Tv control parameter of the cascade controller.
IN_PAR_13
none
Adjusted maximum internal temperature of the cascade
controller.
IN_PAR_14
none
Adjusted minimum internal temperature of the cascade
controller.
IN_PAR_15
none
Band limit (upper)
IN_PAR_16
none
Band limit (lower)