Product description
12
5.3 Technical data
The table contains the technical data of the unit. It refers to the rated voltage
and rated frequency as well as to an ambient temperature of 20°C.
In accordance with IEC 61010-1, the device is designed for safe operation under
the following ambient conditions:
• Indoor use
• Altitude up to 2000 m above sea level
• Ambient tempe5 … +35 °C
• Maximum relative humidity 80% for temperatures up to 31 °C, decreasing
linearly down to 50% relative humidity at 35 °C
• Mains voltage fluctuations up to ±10 % of the nominal voltage permissible
if not otherwise specified
Protection class according to EN 60 529:
• Protection class IP21
Technical data
Working temperature range
°C
+20 … +40
Cooling capacity depends on the temperature
difference between the runback and ambient
temperature
°C
20
15
10
5
Stage 0 W
400
320
220
120
Stage 1 W
550
440
300
180
Circulating pump
Flow rate, max.
Pressure, max.
l/min
bar
3.5
0.49
Noise level at 1 m distance
db (A)
55
Storage tank filling volume
l
0.9
Dimensions (WxHxD)
cm
20x34x30
Protection class according to IEC 60529
IP 21
Permissible ambient temperature range
°C
5 … 40
Weight
kg
11
Mains connection
V/Hz
100 … 230/50/60
Current consumption (230 V)
A
1
Power cable 3x1 mm², 2.5 m
V/A
250/10