Maintenance
17
9 Maintenance
9.1 Replace detachable power cord
The device is equipped with a detachable power cord.
If the power cord needs to be replaced, ensure that the new one is at least
dimensioned for the device power requirements. Insufficiently dimensioned
power cords must not be used. See type plate for mains voltage and current
value.
We recommend only using original JULABO spare parts.
9.2 Emptying
The device must be completely drained if it is to be sent in for technical service
or is to be properly disposed of.
In general, the device should be completely emptied before longer shutdowns or
when there is a change to the external application.
The unit is switched off.
The external application is disconnected from the unit.
1. Disconnect the mains plug of the unit.
2. Remove the cover of the cooling water reservoir.
3. Provide a sufficiently large collecting vessel.
4. Tilt the unit quickly forwards over the collecting vessel.
The water flows from the cooling water reservoir into the collection vessel
provided.
5. When the cooling water reservoir is completely empty, close it with the
cover.
The unit is emptied.