JTECH Medical Commander MN084 Скачать руководство пользователя страница 18

 3 .  Ensure the starting point for breaking action is at the same range of motion 

position for each test and repetition . This is necessary because of the wide 

variations in force the prime movers can oppose depending on the starting 

point of breaking action . 

 4 .  Determine the plane of motion of the segment and ensure that the Muscle 

Tester dynamometer is not sideways or off axis to this motion . Additionally, 

apply force in the direction of motion .

 5 .  Always build up resistance smoothly until the position is broken . In general the 

break should occur within 2-3 seconds . Prolonging the time to break may 

introduce fatigue . Shorter breaks are at higher risk of re-injury or may not allow 

for buildup of maximum contraction . For difficult to break segments, two hands 

may be required .

Grading Muscle Tests

Traditionally, muscle testing uses a grading system for documentation and 

standardization . In the traditional system muscle test grades range from 4/5 “Good/

Normal” to 0/1 “None/Slight Contraction .” With the introduction of the Muscle 

Tester manual muscle tester and the JTECH’s IsoTrack isometric testing station, the 

system has been modified slightly to encompass the additional diagnostic data 

available .

When Should a Muscle Test Be Graded?

Using the traditional subjective method of muscle testing, every test needed to be 

graded just for the purpose of documentation . With Commander’s Muscle Tester 

muscle testing systems, the actual force values are automatically recorded, so the 

need to grade every test is greatly reduced . As a result, tests conducted for 

purposes like initial screenings and monitoring progress are not required to be 

graded . In addition, tests not graded during the examination, or immediately after, 

can be graded at a later date if necessary using the test values recorded on the 

worksheet .

According to the AMA 

Guides to the Evaluation of Permanent Impairment

, a muscle 

test must be graded if it is to be used as part of an impairment rating . Tests where 

the patient cannot exert force against the dynamometer or where the use of a force 

gauge is infeasible should also be graded to provide documentation of the test . 

Grading with the Muscle Testing Dynamometer

Grade 4 or 5:
These grades are assigned if the patient can provide slight to full resistance against 

the dynamometer . 

15

Содержание Commander MN084

Страница 1: ...2016 JTECH Medical MN084_D Muscle Tester...

Страница 2: ...shed by JTECH Medical 7633 S Main St Midvale UT 84047 United States of America 385 695 5000 800 985 8324 www jtechmedical com Authorized Representative for EU Atlantico Systems 34 Oldfield Kingston Ga...

Страница 3: ...est and Test Alt Modes 7 Testing Methods 8 Advanced Console Features 10 Reviewing an Exam 10 Statistics Definitions 12 Deleting an Exam 12 Reports 12 The Muscle Tester Dynamometer Care and Cleaning 13...

Страница 4: ...ANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCES SHALL JTECH BE LIABLE FOR ANY DIRECT INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES THE REMEDIES SET FORTH IN THIS WARR...

Страница 5: ...h off center force accuracy from 94 96 Call JTECH for accuracy report details Includes both round and curved test pads Commander Console Easy to read 2 diagonal LCD Operates during exams either as a w...

Страница 6: ...follow these policies will void the warranty Calibration To ensure accuracy the Muscle Tester and Commander console have been factory calibrated with weights traceable to the National Institute of Sta...

Страница 7: ...onal EasyDocs Impairment Module the system automatically calculates motor deficit impairments from muscle testing results For more information on upgrades contact your JTECH sales representative or ca...

Страница 8: ...n to turn the console off Exams stored in memory are not erased when power is turned off Holding the ON OFF button down for two seconds clears all stored exams The console shuts off automatically afte...

Страница 9: ...less you change the default Default settings do not change when the unit is turned off or when an exam is deleted Defaults can only be reset manually Setting Defaults Three buttons are used when setti...

Страница 10: ...the dynamometer Test recording begins when the default threshold force is exceeded A test ends when force on the dynamometer drops back below the threshold The display then indexes to the next repetit...

Страница 11: ...REP 1 RIGHT NOTE After completing a test Commander prevents inadvertent deletion of tests previously stored in memory by indexing to the next test not completed For example if test 2 was completed pre...

Страница 12: ...rt In this example the test number on the display corresponds to the pre printed test number on the worksheet This method eliminates the necessity of writing the test number on the worksheet before or...

Страница 13: ...tition press TEST CONT To erase the repetition and start over press REDO Deleting a test Press REVIEW then press NEXT until the correct test number is displayed While holding the REVIEW button down pr...

Страница 14: ...e Maximum Force Repetition 2 Left Side CV Left Side Bilateral Deficit Left Side REVIEW REVIEW REVIEW REVIEW REVIEW REVIEW After all statistics have been selected the dis play circles back to the begin...

Страница 15: ...econds or until the unit reactivates When the button is pressed unit console will beep and turn off indicating data has been erased Note All test default values will be retained Reports Muscle testing...

Страница 16: ...humb and index finger at the notches in the dynamometer so your fingers can pull against the underside of the pad Pull the pad straight out from the dynamometer To install a testing pad insert the met...

Страница 17: ...ral ner vous system are ruled out Manual muscle testing should never be used alone to diagnose but should be used in conjunction with other diagnostic testing tools such as the Dualer dual inclinometr...

Страница 18: ...tester and the JTECH s IsoTrack isometric testing station the system has been modified slightly to encompass the additional diagnostic data available When Should a Muscle Test Be Graded Using the trad...

Страница 19: ...ustifying treatment or determining the injured dis eased nerves Grading without the Muscle Testing Dynamometer If the subject does not have sufficient strength to resist against the force gauge or a f...

Отзывы: