World practice tends to refer to points as a combination of letters and numbers (e.g., AP 100 XVI - heart, or AP (XVI) 100 -
heart). This means that the auricular point (AP) under number 100 is located in the XVIth zone of the auricle (concha zone)
and serves to indicate the condition of the cardiovascular system; this zone also serves as the active point for correction of
this state. There are a total of some 200 APs on the surface of the auricle.
This publication does not intend to inform the reader about all APs; its function is to teach you to determine the location of
APs which are most often used in diagnostics and
See chapters
.
Preparation for testing
For two days before the examination the patient should discontinue taking any tonic preparations, drinks or medications. It
is also recommended that the patient does not drink or eat for two hours before testing. It is suggested that the patient relax
for about 15 minutes in a comfortable atmosphere before the testing commences.
The testing should not be carried out in the premises with sources of high frequency electromagnetic fields (such as cell
phones, pages, microwave ovens, TVs, irons, etc). The patient must remove all jewelry, glasses, and watches. During the
testing the patient should sit or lie down comfortably.
!
NOTE! During the examination the physician must not touch the patient with
both hands at the same time. During the testing the patients arms and legs
must not be crossed or touch each other.
Switch on the device.
Insert the diagnostic electrodes into the corresponding sockets.
To enter the BIOREPER program, press • button.
Let the patient hold the passive electrode in his or her hand.
Determining the Individual testing Voltage
Place the active electrode in the in-tan position on the middle line between the eyebrows on the bridge of the nose.
Press and hold INTENSITY* button. The device will automatically select the testing voltage and show the parameter in the
lower right-hand corner of the screen. The lower left-hand corner will show the force of the current passing through the in-
tan point.
When the force of the current passing through the in-tan point will reach 10 mkA, the upper line of the screen will show the
testing voltage as UT=... This will constitute the individual voltage for the testing of the patient.
The device is ready for the diagnostic procedure, emitting the current of the specified power value.
!
NOTE! Should the current ever reach the value of 10 mkA, the screen will
read LITTLE CURRENT, which will mean thatthatthe in-tan point has been
missed, and the active electrode must be repositioned.
BIOREPER testing Program
The patient must hold the passive electrode in his or her palm on the side of the auricle being tested. The active electrode is