Vital Organs

Product Description

1.  Probe

2.  START Button

3.        Button

4.       /      Button

5.  Battery Cover 

Forehead Thermometer Advantages

Infrared Forehead Thermometer measures core body temperature, 

which is the temperature of a body's vital organs. (See Figure 1)  

This thermometer is designed to measure the temperature of the 

skin surface over the temporal artery, a major artery of the head.

The temporal artery is connected to the heart via the carotid artery,

directly leading from the aorta, the main trunk of the arterial system.

It offers constant blood flow. Therefore, body temperature changes

are reflected sooner in the forehead than they are in other parts of 

the body such as oral, rectal and underarm.

3

6

Figure 1

1

2

3

5

4

Real Time Clock

Forehead Mode

Object Mode

The object  mode shows the actual, unadjusted surface  

temperatures

, which is different from the body temperature. It can  

help

 

you

 

to 

monitor if the object temperature is suitable for the baby 

 

or

 

patient

, for example the baby's milk.

 Please see the 

 section to learn how to 

measure the object temperature.

Illustration For Use

The real time clock will be recorded with the memory function and

help you to recognize each measurement result.

 Please see the Real time clock setting section to learn how to 

setup the time in the first use.

The thermometer has been designed for practical use. It's not 

meant

 to replace a visit to the doctor. Please also remember to 

compare 

the measurement result to your regular body temperature.

 Please see the 

 section to learn how to 

measure the body temperature.

Illustration For Use

Memory Mode

There are each 10 sets memories for forehead and object 

measurements. Each memory also records the measurement 

date/time/mode icon.

Please see the 

Selecting A Temperature Scale

 section to learn 

how to change between Celsius and Fahrenheit.

Basic Functions

LCD Display Introduction

4

5

/   Switch

℃ ℉

AM and PM

Low voltage sign

Forehead mode

Temperature Indicator

Unit sign: 

℃ 

or 

Temperature result

Time and Date

sign in Time mode

 Memory mode

Object mode

Beep Alarm

If thermometer detects a body temperature ≥ 37.8℃

100

.

0

under 

forehead

 mode, there will be a long beep sound followed by

three 

s

hort beep sound to warn the user.

(

Distance Detection

If the object is too far away from the probe, the temperature will not 

be measured until the probe is moved within the measured range.

Содержание DET-306b

Страница 1: ...itute for consultation with your physician 4 some parts are small enough to be swallowed 5 Never immerse this device in water or other liquids not waterproof 6 Do not modify this equipment without authorization of manufacturer 7 Do not expose the thermometer to temperature extremes below 25 13 or over 55 131 nor excessive humidity 95 RH 8 Keep the battery away from children 1 There is no gender or...

Страница 2: ... For Use The real time clock will be recorded with the memory function and help you to recognize each measurement result Please see the Real time clock setting section to learn how to setup the time in the first use The thermometer has been designed for practical use It s not meant to replace a visit to the doctor Please also remember to compare the measurement result to your regular body temperat...

Страница 3: ...l This is the only way to accurately diagnose a fever To determine normal temperature take multiple readings when healthy Re measure with a standard digital thermometer for confirmation 2 Users 3 Oils or cosmetics on the forehead may give a lower temperature reading than the actual one must be inside for 30 minutes before taking a measurement Note Users and the thermometer should be in the same am...

Страница 4: ... a reading equal to or higher than 37 8 100 0 In Object mode The display will only be lighted GREEN for 3 seconds when the unit is ready for measurement and a measurement is completed 1 Press the BUTTON The display is activated to show all segments 3 Read the temperature on the display 4 Press the BUTTON to turn off After self checking Figure 2 appears on the display screen with beeps so you can s...

Страница 5: ...to the EMC information Do not put the thermometer into water directly 7 Put the thermometer back to the original packaging after using 15 Match your with a Smart Device 1 If this is your first time using it bind first Open SETTING menu choose Bind and unbind device and select the appropriate model The date and time on your thermometer will automatically be updated when it s connected with your pho...

Страница 6: ...ry cover down as shown in Figure 10 3 Remove battery and install 2 new AAA alkaline batteries as shown in Figure 11 4 Slide battery cover back on Figure 9 Figure 10 Figure 11 Specifications Forehead mode 34 0 43 0 93 2 109 4 Object mode 0 100 32 212 Forehead mode 0 2 0 4 during 35 5 42 0 95 9 107 6 at 15 35 59 0 95 0 operating temperature range 0 3 0 5 for other measuring and operating temperature...

Страница 7: ...vice to the dealers or manufacturer The above recommendations do not supersede the legal requirements The user must always comply with legal requirements for the control of the measurement functionality and accuracy of the device which are required by the scope of relevant laws directives or ordinances where the device is used A clinical summary and procedures for checking calibration are availabl...

Страница 8: ...e Use of the unit in conjunction with non approved accessories can affect the device negatively and alter the electromagnetic compatibility The device should not be used directly adjacent to or between other electrical equipment Electromagnetic Compatibility Information This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cau...

Страница 9: ...d in the frequency range 80 MHz to 2 7 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause interference if it is inadvertently brought into patient areas For this reason an additional factor of 10 3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges c Field stre...

Страница 10: ...or radio control products If you have to do so the device should be observed to verify normal operation The use of accessories and power cord other than those specified with the exception of cables sold by the manufacturer of the equipment or system as replacement parts for internal components may result in increased emissions or decreased immunity of the equipment or system ...

Отзывы: