118
119
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that it is used in such an environment.
Immuni-
ty test
EC
60601
Test
level
Com-
pliance
level
Electromagnetic environment - guidance
Con-
ducted
RF IEC
61000-
4-6
150
kHz to
80
MHz:
3 Vrms
6Vrms
(in ISM
and
ama-
teur
radio
bands)
80%
Am at
1kHz
150 kHz to
80 MHz:
3 Vrms
6Vrms (in
ISM and
amateur
radio
bands)
80% Am at
1kHz
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation appropriate for
the frequency of the transmitter.
Recommended separation distances:
d=0.35
P
;
d=1.2
P