ATTACHING CIRCUITS
F-CIRCUIT ATTACHMENT
For use with animals that weigh
approximately 15 lbs. (7kg) or more.
u
Connect the longer, single limb of the circuit to
the inhalation valve.
u
Connect the shorter, corrugated tube to the
exhalation valve.
u
Connect a rebreathing bag (appropriate to the
patient size) to the port located below the pop-
off valve.
u
Flip the Rebreathing/Non-rebreathing switch to
Rebreathing when using this circuit.
Note: It is
not necessary to remove the non-rebreathing
circuit.
NON-REBREATHING CIRCUIT ATTACHMENT
For use with animals that weigh
approximately 15 lbs. (7kg) or less.
u
Insert the adapter on the end of the long, clear
tubing into the Common Gas Outlet (located
above the Rebreathing/Non-rebreathing switch).
u
Attach the corrugated tubing to the waste gas
scavenging system.
u
Confirm the bag bleed valve/slide valve is open
on the non-rebreathing circuit.
u
Flip the switch to non-rebreathing when using
the non-rebreathing circuit. It is not necessary to
remove the rebreathing circuit.
WARNING
u
As a patient safety feature, the oxygen flush
will not work with a non-rebreathing circuit. The
oxygen flush is only piped to the rebreathing
absorber.