Jorvet J0558X Скачать руководство пользователя страница 9

9

ATTACHING CIRCUITS

F-CIRCUIT ATTACHMENT

For use with animals that weigh 

approximately 15 lbs. (7kg) or more.

u

 

Connect the longer, single limb of the circuit to 
the inhalation valve.

u

 

Connect the shorter, corrugated tube to the 
exhalation valve.

u

 

Connect a rebreathing bag (appropriate to the 
patient size) to the port located below the pop-
off valve.

u

 

Flip the Rebreathing/Non-rebreathing switch to 
Rebreathing when using this circuit. 

Note: It is 

not necessary to remove the non-rebreathing 
circuit.

NON-REBREATHING CIRCUIT ATTACHMENT

For use with animals that weigh 

approximately 15 lbs. (7kg) or less.

u

 

Insert the adapter on the end of the long, clear 
tubing into the Common Gas Outlet (located 
above the Rebreathing/Non-rebreathing switch).

u

 

Attach the corrugated tubing to the waste gas 
scavenging system.

u

 

Confirm the bag bleed valve/slide valve is open 
on the non-rebreathing circuit.

u

 

Flip the switch to non-rebreathing when using 
the non-rebreathing circuit. It is not necessary to 
remove the rebreathing circuit.

WARNING

u

 

As  a  patient  safety  feature,  the  oxygen  flush 
will not work with a non-rebreathing circuit. The 
oxygen  flush  is  only  piped  to  the  rebreathing 
absorber.

Содержание J0558X

Страница 1: ...TOP LINE ANESTHESIA SYSTEM J0558X J0558Z USER MANUAL The JorVet Top Line Anesthesia System pictured above shows optional accessories J0558Y Bain Block and J0744SV Concentrator...

Страница 2: ...e Maintenance Troubleshooting and Specifications Warranty and Service 03 04 05 06 07 08 09 10 11 12 13 Thank you for using our JorVet J0558X and J0558Z Anesthesia System We want you to get the most ou...

Страница 3: ...3 TOP LINE ANESTHESIA SYSTEM MODELS J0558X J0558Z Shown with optional J0558Y Bain Block and J0744SV Concentrator IMPORTANT VAPORIZER NOT INCLUDED FOR BOTH MODELS...

Страница 4: ...helf to provide more workspace and keep the patient s vitals and anesthesia controls in one area JorVet Top Line Anesthesia System Assembly u H Frame Stand Mounting Plate Assembly and Rebreathing Head...

Страница 5: ...en flush will not work with a non rebreathing circuit The oxygen flush is only piped to the rebreathing absorber u Always operate inhalant anesthetic equipment with waste gas scavenging systems Failur...

Страница 6: ...witch Rebreathing Non rebreathing 9 10 Pop off APL Valve 11 Rebreathing Bag Port Absorber Canister 12 13 Absorber Canister Knob 14 Manometer 15 16 Oxygen Flowmeter Flush Valve 1 2 Accessory Mounting A...

Страница 7: ...r block positioned on the front face of the mounting plate and tighten Note Two people recommended for this step Mounting hardware supplied with vaporizer u Connect the inlet and outlet endcaps using...

Страница 8: ...essure small green tubes from the anesthesia machine to the E tank regulator by pushing the tubing into the fittings Once resistance is felt continue to push the tubing slightly further to secure conn...

Страница 9: ...g circuit NON REBREATHING CIRCUIT ATTACHMENT For use with animals that weigh approximately 15 lbs 7kg or less u Insert the adapter on the end of the long clear tubing into the Common Gas Outlet locate...

Страница 10: ...N CARRIER GAS SUPPLY REGULATOR Compressed Tank in house or Concentrator Not needed for Concentrator Regulated Oxygen Carrier Gas Hose Anesthetic Agent and Oxygen Carrier Gas Mixture Flowmeter 0 4 LPM...

Страница 11: ...tton or turn on oxygen flow meter until the bag is fully distended and or manometer reads 30 40 cm of H2O pressure Re open the pop off valve to your standard setting Squeeze the bag with your hand sti...

Страница 12: ...gas system function attenuation 4 Confirm vaporizer has an adequate amount of agent 1 Tighten any loose connections vaporizer endcaps tubing connections 2 Replace breathing circuit or bag Fix loose co...

Страница 13: ...maged by neglect or accident tampered with not properly maintained not installed in strict compliance with applicable codes and ordinances or repaired by persons not authorized by Jorgensen Laboratori...

Страница 14: ...Jorgensen Laboratories Inc Loveland CO 80538 800 525 5614 fax 970 663 5042 www jorvet com info jorvet com Professional Veterinary Equipment Supplier...

Отзывы: