CONCENTRATOR KIT INSTALLATION – MODEL J0558X
Stud Hardware Installed on Base Plate
u
4 – 3/8”-24 x 2” Hex Bolt
u
8 – 3/8” x 24 Hex Nut
u
12 – 3/8” SS Washer
u
4 – 3/8” Cushioning Washer
In order as installed: Bolt, SS Washer, Base Plate, SS Washer, 2x Hex Nut, SS Washer,
Cushioning Washer.
FOLLOW ALL STEPS ACCORDINGLY
u
Unpack concentrator and remove wheels by pulling away from (base) of concentrator.
u
Install stud hardware on base plate in correct holes based on where wheels were
located on concentrator.
u
Set concentrator on studs sticking up from base plate.
Note: Two people recommended
for this step.
u
The single E-cylinder manifold will be pre-installed on the middle support shelf. Ensure
the brass and black washer/seal is in place prior to installing the e-cylinder.
u
Attach the two high pressure, small green tubes from the anesthesia machine to the
E-tank regulator by pushing the tubing into the fittings. Once resistance is felt, continue
to push the tubing slightly further to secure connection. Gently pull on tubing to test for
proper connection.
Note: The E-cylinder must be turned on to operate the oxygen
flush.
u
Attach low pressure, large green tubing from the anesthesia machine to the concentrator
by threading the green female fitting onto the male fitting on the concentrator.
u
Optional: Use supplied cable tie to organize the green tubing.
WARNING
u
When using the oxygen concentrator, flip the switch on the front of the anesthesia
machine to concentrator prior to turning the concentrator on.
u
When using the oxygen concentrator, turn anesthesia machine flowmeter on to allow
flow. The concentrator alarm will sound if this step is not followed.
NOTE
u
When using the oxygen concentrator, turn the concentrator flow to match the flowrate
on the anesthesia machine.