PRODUCT OPERATION

REFER TO PRODUCT INSTALLATION GUIDE FOR ILLUSTRATION

FOLLOW ALL STEPS ACCORDINGLY

u

 

Remove all cartons from the main package.

u

 

Remove the machine H-Frame base unit.

u

 

Unpack and inspect remaining components for damage.  Be sure to retain the mounting 
hardware contained inside package.

u

 

Attach middle shelf to center post using 3 bolts.

u

 

Attach center post with middle shelf attached to H-frame base with 4 bolts.

u

 

Attach mounting plate assembly to center pole with 4 bolts. 

Note: Two people 

recommended for this step.

u

 

Attach rebreathing head to mounting plate assembly using supplied bolts and allen 
wrench.

u

 

Connect black supply line to the hose barb on the left side of the rebreathing head.

 

Warning: Open pop-off valve prior to use.

u

 

Install anesthetic vaporizer onto frame with vaporizer spacer block positioned on the 
front face of the mounting plate and tighten. 

Note: Two people recommended for this 

step. Mounting hardware supplied with vaporizer.

u

 

Connect the inlet and outlet endcaps using a push and twist action to provide a leak 
free attachment. 

Note: This step can be performed prior to installing the vaporizer 

onto the frame if needed.

u

 

Connect oxygen (carrier gas) supply source to the DISS Male fitting on the back of the 
flowmeter. 

Note: For Model J0558X see next page for Concentrator Kit installation.

u

 

Turn the oxygen (carrier gas) supply on.

u

 

Connect appropriate active or passive waste gas scavenging system to the 19mm 
outlet of the pop-off valve.

u

 

Ensure vaporizer has been filled per manufacturer’s instructions and is ready for use.

u

 

Fill absorber canister to line indicated on canister. Use supplied downtube cover (on 
chain) when filling. Attach absorber canister to the rebreathing head and tighten.

 

Note: Absorbent not included.

u

 

Leak test following instructions on 10 second pressure test card.

Содержание J0558X

Страница 1: ...TOP LINE ANESTHESIA SYSTEM J0558X J0558Z USER MANUAL The JorVet Top Line Anesthesia System pictured above shows optional accessories J0558Y Bain Block and J0744SV Concentrator...

Страница 2: ...e Maintenance Troubleshooting and Specifications Warranty and Service 03 04 05 06 07 08 09 10 11 12 13 Thank you for using our JorVet J0558X and J0558Z Anesthesia System We want you to get the most ou...

Страница 3: ...3 TOP LINE ANESTHESIA SYSTEM MODELS J0558X J0558Z Shown with optional J0558Y Bain Block and J0744SV Concentrator IMPORTANT VAPORIZER NOT INCLUDED FOR BOTH MODELS...

Страница 4: ...helf to provide more workspace and keep the patient s vitals and anesthesia controls in one area JorVet Top Line Anesthesia System Assembly u H Frame Stand Mounting Plate Assembly and Rebreathing Head...

Страница 5: ...en flush will not work with a non rebreathing circuit The oxygen flush is only piped to the rebreathing absorber u Always operate inhalant anesthetic equipment with waste gas scavenging systems Failur...

Страница 6: ...witch Rebreathing Non rebreathing 9 10 Pop off APL Valve 11 Rebreathing Bag Port Absorber Canister 12 13 Absorber Canister Knob 14 Manometer 15 16 Oxygen Flowmeter Flush Valve 1 2 Accessory Mounting A...

Страница 7: ...r block positioned on the front face of the mounting plate and tighten Note Two people recommended for this step Mounting hardware supplied with vaporizer u Connect the inlet and outlet endcaps using...

Страница 8: ...essure small green tubes from the anesthesia machine to the E tank regulator by pushing the tubing into the fittings Once resistance is felt continue to push the tubing slightly further to secure conn...

Страница 9: ...g circuit NON REBREATHING CIRCUIT ATTACHMENT For use with animals that weigh approximately 15 lbs 7kg or less u Insert the adapter on the end of the long clear tubing into the Common Gas Outlet locate...

Страница 10: ...N CARRIER GAS SUPPLY REGULATOR Compressed Tank in house or Concentrator Not needed for Concentrator Regulated Oxygen Carrier Gas Hose Anesthetic Agent and Oxygen Carrier Gas Mixture Flowmeter 0 4 LPM...

Страница 11: ...tton or turn on oxygen flow meter until the bag is fully distended and or manometer reads 30 40 cm of H2O pressure Re open the pop off valve to your standard setting Squeeze the bag with your hand sti...

Страница 12: ...gas system function attenuation 4 Confirm vaporizer has an adequate amount of agent 1 Tighten any loose connections vaporizer endcaps tubing connections 2 Replace breathing circuit or bag Fix loose co...

Страница 13: ...maged by neglect or accident tampered with not properly maintained not installed in strict compliance with applicable codes and ordinances or repaired by persons not authorized by Jorgensen Laboratori...

Страница 14: ...Jorgensen Laboratories Inc Loveland CO 80538 800 525 5614 fax 970 663 5042 www jorvet com info jorvet com Professional Veterinary Equipment Supplier...

Отзывы: