CIRCUIT DESCRIPTIONS, WARNINGS, AND NOTES

The F-Circuit is a rebreathing circuit for use

with animals that weigh approximately 15 lbs. (7kg) or more.

The Non-rebreathing Circuit is a circuit for use

with animals that weigh approximately 15 lbs. (7kg) or less.

WARNINGS

u

 

Operator must read entire manual prior to operation and have proper training in 
the use of inhalant anesthetic equipment. Failure to do so may harm the animal.

u

 

Do not connect unregulated compressed carrier gas to the Top Line Anesthesia 
System. Compressed oxygen, or another carrier gas, should be supplied at 50-55 psi.

u

 

Pop-off (APL) valve is closed for shipping. Completely open the pop-off valve prior to 
use. The pop-off valve should be kept open during normal operation and only closed 
when performing a leak check or using a ventilator.

u

 

As a patient safety feature, the oxygen flush will not work with a non-rebreathing circuit. 
The oxygen flush is only piped to the rebreathing absorber.

u

 

Always operate inhalant anesthetic equipment with waste gas scavenging systems. 
Failure to control waste anesthetic gas may cause harmful exposure to personnel.

u

 

When  using  the  oxygen  concentrator,  turn  the  anesthesia  machine  flowmeter  on  to 
allow flow. The concentrator alarm will sound if this step is not followed.

NOTE

u

 

When using the oxygen concentrator with the J0558X model, the oxygen flush will not 
operate without the E-cylinder turned on. The oxygen flush is piped to the e-cylinder 
only.

u

 

When using the oxygen concentrator, turn the concentrator flow to match the flowrate 
on the anesthesia machine.

u

 

Do not overtighten the flowmeter knob. When the float is resting at the bottom of the 
flowmeter, the flowmeter is off. Overtightening will damage the needle valve and result 
in leakage.

u

 

Anesthetic Vaporizer and carrier gas (oxygen) accessories are not included. Please 
follow the installation guides for these products.

Содержание J0558X

Страница 1: ...TOP LINE ANESTHESIA SYSTEM J0558X J0558Z USER MANUAL The JorVet Top Line Anesthesia System pictured above shows optional accessories J0558Y Bain Block and J0744SV Concentrator...

Страница 2: ...e Maintenance Troubleshooting and Specifications Warranty and Service 03 04 05 06 07 08 09 10 11 12 13 Thank you for using our JorVet J0558X and J0558Z Anesthesia System We want you to get the most ou...

Страница 3: ...3 TOP LINE ANESTHESIA SYSTEM MODELS J0558X J0558Z Shown with optional J0558Y Bain Block and J0744SV Concentrator IMPORTANT VAPORIZER NOT INCLUDED FOR BOTH MODELS...

Страница 4: ...helf to provide more workspace and keep the patient s vitals and anesthesia controls in one area JorVet Top Line Anesthesia System Assembly u H Frame Stand Mounting Plate Assembly and Rebreathing Head...

Страница 5: ...en flush will not work with a non rebreathing circuit The oxygen flush is only piped to the rebreathing absorber u Always operate inhalant anesthetic equipment with waste gas scavenging systems Failur...

Страница 6: ...witch Rebreathing Non rebreathing 9 10 Pop off APL Valve 11 Rebreathing Bag Port Absorber Canister 12 13 Absorber Canister Knob 14 Manometer 15 16 Oxygen Flowmeter Flush Valve 1 2 Accessory Mounting A...

Страница 7: ...r block positioned on the front face of the mounting plate and tighten Note Two people recommended for this step Mounting hardware supplied with vaporizer u Connect the inlet and outlet endcaps using...

Страница 8: ...essure small green tubes from the anesthesia machine to the E tank regulator by pushing the tubing into the fittings Once resistance is felt continue to push the tubing slightly further to secure conn...

Страница 9: ...g circuit NON REBREATHING CIRCUIT ATTACHMENT For use with animals that weigh approximately 15 lbs 7kg or less u Insert the adapter on the end of the long clear tubing into the Common Gas Outlet locate...

Страница 10: ...N CARRIER GAS SUPPLY REGULATOR Compressed Tank in house or Concentrator Not needed for Concentrator Regulated Oxygen Carrier Gas Hose Anesthetic Agent and Oxygen Carrier Gas Mixture Flowmeter 0 4 LPM...

Страница 11: ...tton or turn on oxygen flow meter until the bag is fully distended and or manometer reads 30 40 cm of H2O pressure Re open the pop off valve to your standard setting Squeeze the bag with your hand sti...

Страница 12: ...gas system function attenuation 4 Confirm vaporizer has an adequate amount of agent 1 Tighten any loose connections vaporizer endcaps tubing connections 2 Replace breathing circuit or bag Fix loose co...

Страница 13: ...maged by neglect or accident tampered with not properly maintained not installed in strict compliance with applicable codes and ordinances or repaired by persons not authorized by Jorgensen Laboratori...

Страница 14: ...Jorgensen Laboratories Inc Loveland CO 80538 800 525 5614 fax 970 663 5042 www jorvet com info jorvet com Professional Veterinary Equipment Supplier...

Отзывы: