7
Chapter 3 - Basic Operation
continued
3.1
Controls,
Indicators, and
Connections
continued
3.2
Instrument
Preparation and
Hand Piece
Attachment
9
Status Lights
Ready Light. This light is illuminated when the system is ready to use. This light is off while the
Hand Piece is energized or in STANDBY mode.
Variable Light. This light is flashed on and off when the Hand Piece is energized at the
variable (VAR) power level.
Full Light. This light is flashed on and off when the Hand Piece is energized at the full
(FULL) power level.
10 Standby Button
This button toggles between STANDBY and READY mode. Push this button once to enter
STANDBY mode. In STANDBY mode, all power is removed from the Hand Piece. The Increase,
Decrease, and Save buttons are disabled as is the Foot Switch. Use the STANDBY mode when
replacing instrument blades.
11 Release Button
Push this button to detach the Hand Piece cord connection from the Cooling Air Source
Connector on the Generator.
12 Cooling Air Source
Refer to the appropriate Hand Piece package insert.
13 Electrical Receptacle
Delivers the energy from the Generator to the Hand Piece.
*
Audio Annunciator
The annunciator beeps to indicate instrument activity or to indicate error conditions. During
normal activation, the annunciator beeps on and off when either foot pedal is pressed. A slower
beep indicates that LEVEL 1 through LEVEL 4 is activated and a faster beep indicates that
LEVEL 5 is activated. If the generator does not energize the instrument, the annunciator will emit
a continuous beep when either pedal is pressed.
Instrument Preparation and Hand Piece Attachment
Step
Instruction
1
Ensure that the Generator Power Switch is off during instrument preparation.
WARNING: To avoid injury, the activatedU
LTRA
C
ISION
H
ARMONIC
S
CALPEL
instrument should not be in contact with drapes or other flammable materials. During
prolonged activation, the instrument blades may become hot. Avoid unintended blade
contact with tissue, drapes, surgical gowns, or other unintended sites after activation.
2
Plug the line cord into a grounded receptacle:
Model GEN01:
115 Vac, 60 Hz
Model GEN32:
230 Vac, 50 Hz
Note: Do Not Connect the Generator To An Ungrounded Outlet.
3
Connect the Foot Switch to the rear panel of the Generator. Align the red dot on the Foot
Switch 4-pin connector with the red dot 4-pin connector on the lower left side of the
Generator back panel.