6.3.4
Low Voltage
Transformer

Chapter 6 - Detailed Diagnostics  

continued

The compressor is screwed to the base plate on rubber feet in the back portion of the Generator. It runs on line
voltage supplied through the fusing and distribution module. When the front panel switch is turned on, the cooling
fins on the compressor quickly become hot and the compressor vibrates; this is normal and is not cause for
replacement. For a functional description see section 2.2.

With power on, remove the Hand Piece air connector. A steady flow of air should be detectable. If available, a
pressure gauge can be used to measure the air pressure, which should be greater than or equal to 4.5 pounds per
square inch. Reconnect the hand piece air connection or plug.

If power is available to the compressor and there is no or reduced air flow, check the hose leading from the
compressor for any perforations or restrictions. Also check the plastic filter housing and filter, located on the side 
of the compressor, for any blockage before replacing the compressor (models HP050, HP051, H1TUV, HPTUV).

NOTE: The compressor filter, located in the plastic housing, should be removed and cleaned at least once every
two years. To remove the filter, unscrew the wing nut securing the plastic housing to the compressor.

Replacement:

1.

Disconnect the line cord and unscrew the four phillips head machine screws from the base plate. 

2.

Disconnect the two primary wires from the fusing module and remove the grounding wire from the 
base plate. 

3.

With pliers, unscrew the hose fitting. Remove the compressor and replace with the appropriate part listed 
in Appendix C.

The low voltage transformer is screwed to the left side of the base plate underneath the generator PCB and receives
line voltage from the fusing and distribution module. For a functional description see section 2.2.

The output of this transformer terminates at J1 on the generator PCB. Proceed through the following steps to test 
for the proper output voltage:

1.

Remove the power cord. Remove the plug at J1 and clip the digital voltmeter to
locations 3 (white wire) and 4 (green wire). 
NOTE: Location 1 is next to the filled keying location.

WarningStay clear of locations 7 and 8; they are high voltage (250 to 260 VAC).

2.

Supply power and turn the front panel switch on. With nominal line voltage, the open
circuit voltage between locations 3 and 4 should be between 33 and 35 volts AC.

3.

Repeat this process and verify the voltage between locations 2 (yellow wire) and
4 (green wire) is between 16 and 17 volts AC.

4.

Repeat this process and verify the voltage between locations 5 (red wire) and 
6 (black wire) is between 9.5 and 10.5 volts AC.

5.

Remove the power cord and replace the J1 plug.

If power is available to the transformer and there is no output voltage, remove the line cord, refer to the wiring
diagram (Appendix B), and inspect the wiring leading to the transformer. If the wiring appears intact, replace the
transformer.

20

6.3.3
Compressor

Содержание Ethicon Endo-Surgery ULTRACISION HARMONIC SCALPEL

Страница 1: ...ULTRACISION HARMONIC SCALPEL Service Manual...

Страница 2: ...e Shooting Guide 5 2 Diagnostic Software 5 3 Extended Functional Test Chapter 6 Detailed Diagnostics 6 1 Test Equipment 6 2 Hand Piece Assembly 6 3 Generator and Foot Switch 6 3 1 Power Entry Module 6...

Страница 3: ...matic 115 Volt Wir ing Model GEN01 230 Volt Wiring Models GEN22 GEN32 115 Volt Routing Model GEN01 230 Volt Routing Models GEN22 GEN32 Generator PCB Schematic Generator PCB Component Location Keyboard...

Страница 4: ...ustic drive train is formed The acoustic system or acoustic drive train consists of three components Acoustic Transducer converts the electrical energy into motion The transducer consists of four piez...

Страница 5: ...secondary function of the Generator PCB is to provide a central point of connectivity for the other PCBs and modules in the Generator The third function is to rectify filter and regulate the low volta...

Страница 6: ...ENTRY MODULE FILTER FUSING AC DISTRIBUTION HIGH VOLTAGE TRANSFORMER MICROPROCESSOR PCB GENERATOR PCB COMPRESSOR AIR models HP050 HP051 HPTUV H1TUV KEYBOARD INTERLOCK FOOT SWITCH HAND PIECE 55 KHz AC L...

Страница 7: ...NDBY mode When the Generator power is first turned on this display briefly indicates that a SELF TEST is in progress This display is also used for alarm messages When activated by service personnel th...

Страница 8: ...ator to the Hand Piece Audio Annunciator The annunciator beeps to indicate instrument activity or to indicate error conditions During normal activation the annunciator beeps on and off when either foo...

Страница 9: ...w the Hand Piece to dry STANDBY Mode is recommended during this period of time 8 If the Generator is in STANDBY Mode press the Standby Button to return it to READY Mode If the generator is not already...

Страница 10: ...ly selected The line rate may be toggled between 50 Hz and 60 Hz by pressing the Increase Button A validation sequence is required to change the line rate In addition to updating the line rate the mem...

Страница 11: ...the instrument with caution as it may become hot during operation in air Inside of Hand Piece is still warm from steam or Allow Hand Piece to cool at room temperature for 15 30 flash sterilization mi...

Страница 12: ...Hand Piece Blade was attached to the Hand Piece with Push Standby Button Remove and discard the blade extender excessive force from the Hand Piece Attach a new instrument blade to the Hand Piece Mate...

Страница 13: ...nd emitted from Hand Piece Probable cause Corrective Action Instrument blade is not fully seated on Re tourque the instrument blade with an alternate torque Hand Piece blade mount blade wrench and or...

Страница 14: ...3 8 to replace the microprocessor PCB CRC Five digit number followed by the line rate is displayed This indicates that the second page of the service mode has been accessed When the software detects...

Страница 15: ...sible Problems Display voltage or current is low 15 or 40 respectively The high voltage AC is not reaching the generator PCB or the generator PCB s switching power supply is not working Verify that th...

Страница 16: ...any other equipment especially electrosurgical devices If the problem persists replace the generator PCB see section 6 3 6 10 The cabling leading from the generator PCB to the keyboard became disconne...

Страница 17: ...y a continuous tone and LED illumination While still holding the pedal down remove the blade from the cutting material and verify that the blade again starts to resonate This is indicated by the norma...

Страница 18: ...tinuity between pin 1 and the tip of the blade see Figure 3 for pin location For Models HPTUV and H1TUV verify continuity betwen pin 2 and the tip of the blade see Figure 3 for pin location If an open...

Страница 19: ...ION RETURN BLACK 4 5 7 6 5 4 3 2 1 6 7 2 1 3 DOMESTIC HP050 HP051 INTERNATIONAL H2TUV H3TUV Hand Piece Models HP050 HP051 HPTUV H1TUV Hand Piece Models HP052 HP053 H2TUV H3TUV NO CONNECTION NO CONNECT...

Страница 20: ...erators Models GEN22 and GEN32 For a functional description see section 2 2 Verification 1 All fuses used in the generator are time delayed and therefore may appear closed when they are actually open...

Страница 21: ...e base plate 2 Disconnect the two primary wires from the fusing module and remove the grounding wire from the base plate 3 With pliers unscrew the hose fitting Remove the compressor and replace with t...

Страница 22: ...le to the transformer and there is no output voltage remove the line cord refer to the wiring diagram Appendix B and inspect the wiring leading to the transformer If the wiring appears intact replace...

Страница 23: ...the case re test the generator PCB with a substitute microprocessor PCB before replacing this PCB If the pulse at TP16 is not synchronous with the line rate replace the generator PCB 2 Verify that the...

Страница 24: ...lock portion of this PCB is working properly as follows 1 Access the first page of the software diagnostic mode as described in section 5 2 and diagrammed in Appendix A 2 Press the Standby Button twic...

Страница 25: ...Chapter 6 Detailed Diagnostics Figure 4 24 Microprocessor PCB Service Position...

Страница 26: ...cified in Appendix C The LCD is mounted to the back of the front panel assembly It receives control signals and supply voltage from J3 on the generator PCB and backlight power from J10 on the generato...

Страница 27: ...ains E S D protection resistors in the cable connector and must therefore be tested together as a unit as follows 1 Remove the foot switch connector from the back panel of the generator and orient it...

Страница 28: ...ce 7 Hand Piece 6 Cable 10 Cable 10 5 Connectors Electric 7 pin Electric 7 pin Air 1 4 CPC Air None Sterilization Steam EO or STERRAD Steam EO or STERRAD Switch Dual SPST Normally Off Voltage 12 Volts...

Страница 29: ...tches Released STANDBY key STANDBY or UP DOWN service mode No Hand Piece Power READY LED OFF switches UP DOWN SAVE keys disabled 3 Level 3 Left foot or ring switch VAR LED flashes slow beep Level 5 ri...

Страница 30: ...ETHICON ENDO SURGERY INC ETHICON ENDO SURGERY INC 1996 Cincinnati OH 45242 2839 USA P40238P02 BAR CODE P40238P02...

Страница 31: ...99 03869 CB 7 29 99 PRODUCT ARTWORK PACKAGE COMPONENT CODE NUMBER NUMBER GEN01 A83513P00 P40238P02 GEN22 GEN32 CONFIDENTIAL NOT TO BE REPRODUCED OR USED IN ANY WAY WITHOUT WRITTEN ETHICON ENDO SURGERY...

Отзывы: