36
GEN04
Service Manual
Error Code 7: Hand Switch
Error Code 7 indicates the hand switch is stuck in the ON position.
Confirm contacts in the distal end of hand piece and in the proximal
end of the hand switching adaptor are dry or replace the hand
switching adaptor.
Note: If the error persists, replace hand switch.
Fig. 9-7 Hand Activation Assembly Error Screens
Содержание ETHICON ENDO-SURGERY HARMONIC 300
Страница 1: ...HARMONIC Generator 300 System Service Manual ...
Страница 2: ...GEN04 ...
Страница 6: ...4 GEN04 Service Manual ...
Страница 10: ...8 GEN04 Service Manual ...
Страница 16: ...14 GEN04 Service Manual ...
Страница 24: ...22 GEN04 Service Manual ...
Страница 26: ...24 GEN04 Service Manual ...
Страница 28: ...26 GEN04 Service Manual ...
Страница 44: ...42 GEN04 Service Manual ...
Страница 50: ...48 GEN04 Service Manual Power Supply Fig 12 5 Power Supply Troubleshooting Flowchart ...
Страница 51: ...49 GEN04 Chapter 12 Hardware Troubleshooting LCD Display Fig 12 6 LCD Display Troubleshooting Flowchart ...
Страница 52: ...50 GEN04 Service Manual Front Bezel Fig 12 7 Front Bezel Troubleshooting Flowchart ...
Страница 54: ...52 GEN04 Service Manual Unit Short Test Fig 12 9 Unit Short Test Troubleshooting Flowchart Cont ...
Страница 55: ...53 GEN04 Chapter 12 Hardware Troubleshooting Cooling Fan Fig 12 10 Cooling Fan Troubleshooting Flowchart ...
Страница 58: ...56 GEN04 Service Manual ...
Страница 62: ...60 GEN04 Service Manual ...
Страница 66: ...64 GEN04 Service Manual ...
Страница 68: ...66 GEN04 Service Manual ...
Страница 72: ...70 GEN04 Service Manual ...
Страница 76: ...74 GEN04 Service Manual ...
Страница 77: ...75 GEN04 Appendix B GEN04 Main Assembly Drawing ...
Страница 79: ...77 GEN04 Appendix C GEN04 Top Level Diagram Hand Piece ID Hand Piece Transducer ...