66 DePuy Synthes Power Tools Battery Power Line II User’s Manual
Problem
Possible causes
Solution
Drive unit or attachment
becomes excessively hot
Drive unit or attachment is used outside
the specification
Allow drive unit or attachment to cool.
(see Duty cycles on page 68)
The cutting tool is blunt
Replace the cutting tool
Visible physical damage
on items
Battery was accidentally reprocessed
Replace battery. Send damaged battery
to DePuy Synthes service center
Drive unit, attachment, battery casing,
insertion shield was dropped
Replace damaged items. Send damaged
items to DePuy Synthes service center
Battery is faulty
Improper care/maintanance
Replace battery and send the battery to
DePuy Synthes service center
Battery casing jams when
inserting or removing from
drive unit
Coupling mechanism has not been lubricated Clean and lubricate according to care
and maintenance guidelines
Coupling mechanism is damaged
Send damaged item to DePuy Synthes
service center
Battery casing lid is difficult
to open and close
Sealing ring has not been lubricated
Clean and lubricate according to care
and maintenance guidelines
General (continued)
Troubleshooting