76 DePuy Synthes Power Tools Battery Power Line II User’s Manual
Table 4: Immunity
(not life-supporting devices)
Guidance and Manufacturer’s Declaration—
Electromagnetic Immunity
The BPL or BPL II System is intended for use in the
electromagnetic environment specified below. The
customer or the user of the BPL or BPL II System should
ensure that it is used in such an environment.
Immunity Test
Standard
IEC 60601
Test Level
Compliance
Level
Recommended
Separation Distance
c
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
n/a
d = 0.35
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 800 MHz
E1 = 10 V/m
80 MHz to 800 MHz
d = 0.35
80 MHz to 800 MHz
Radiated RF
IEC 61000-4-3
3 V/m
800 MHz to 2.5 GHz
E2 = 10 V/m
80 MHz to 800 MHz
d = 0.7
0.8 GHz to 2.5 GHz
Electromagnetic Environment—Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the BPL or BPL II
System, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Notes
• At 80 MHz and 800 MHz, the higher frequency
range applies.
• These guidelines may not apply in all situations.
Electromagnetic propagation is affected by
absorption and reflection from structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
in the location in which the BPL or BPL II System is used exceeds the applicable
RF compliance level above, the BPL or BPL II System or the device contains the
BPL or BPL II System should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as
re-orienting or relocating the device containing the BPL or BPL II System.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 10 V/m.
c Possible shorter distances outside ISM bands are not considered to have a
better applicability of this table.
Electromagnetic Compatibility
Accompanying Documents According to IEC 60601-1-2, 2007, Clause 6