23
English
Oxford
®
Mermaid & Ranger
16. Maintenance Schedule
Joerns Healthcare Ltd
recommend the following tasks are carried out by authorised service dealers
only.
FLOOR FIXING:
Check the base plate fixing is sound and secure. Follow the test routine as
detailed in the installation instructions. Wooden floors require inspection at more frequent intervals.
MAST TO BASE:
Check the four Nyloc nuts securing the mast to the base plate are fully tightened.
Check also for signs of grease leakage between the mast and the base plate.
CHAIR SUPPORT CLAMP:
Check the chair support clamp is correctly orientated as per the
Setting Up instructions. Make sure the seat does not touch the side or bottom of the bath. Check
the seat locks in the correct positions. CHECK the operation of the seat lock.
SUPPORT ARM:
Check the support arm has the safety screw in position and tightened. Check the
seat is securely attached to the support arm and the fixing nuts and bolts are fully tightened.
SEAT:
Check the seat for splits or damage. Check the Mermaid seat for presence of nine safety
plugs fitted to the seat moulding (Lifts before December 1998). Check that the seat is securely
fitted to the frame.
RANGER BASE CASTORS:
Check all castors are free running. Check for secure fixing to the
base legs. Check brakes for correct operation.
SAFETY LATCHES:
Check the correct function of the safety latches on the Ranger Chair and the
Ranger support arm.
LAP STRAP FITTING:
Check that the lap strap is fitted securely to the seat frame and is not
frayed, worn or damaged.
SEAT FIXING:
Check that the seat is securely fitted to the frame.
HOIST CLEANING:
Refer to section 13.
LAP STRAP CLEANING:
Refer to section 13.
LOAD TEST:
The load test should be carried out in accordance with the manufacturer’s test
procedures and the directive detailed in EN ISO 10535:2006 - Annex B - Periodic Inspection B1 -
see excerpt below. It is strongly recommended that an authorised service dealer carry out the test.
EN ISO 10535:2006
Annex B
(Informative)
Periodic Inspection
B. 1
Periodic inspection of the hoist should be undertaken at the time intervals stated by the
manufacturer, but at least once a year. By periodic inspection is meant a visual examination
(particularly of the hoist’s load bearing structure and lifting mechanism with attachments, brakes,
controls, safety devices and person-support devices) and whatever function tests and maintenance
measures may be required, e.g. adjustment of brakes, tightening of fasteners.
Every inspection should include a working load test of one (1) lifting cycle with the maximum load.
CERTIFICATION:
An authorised service dealer will issue a test certificate after satisfactory
completion of the load test. This certificate will be valid for the period stated on the certificate.