Joerns Support Surface
DermaFloat LR Model
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
Turn Angle
The
Turn Angle
feature allows for the desired turn
angle to be selected. The mattress can make a half or
full turn based on the optimal therapy for
the resident.
Additional Features
Warning:
A possible fire hazard exists. This product is
suitable for use with oxygen administering equipment
of the nasal, mask, or half bed-length, tent-type only.
To prevent personal injury or equipment damage,
ensure that the oxygen tent does not extend below
the mattress.
CPR
The hose connection at the control unit is marked
CPR
. Disconnect the hose from the control unit.
Deflation times will vary based on resident weight and
profile. To resume therapy, reattach the hose to the
control unit.
Transport
To transport the resident in bed, turn the control unit
off. Unplug the power cord from the outlet. Do not
disconnect the hose connection at the control unit.
The lower support cell will prevent the resident from
bottoming out for up to 12 hours.
Power Failure
If power fails to the system an alarm will sound and
the
Power Fail
LED will turn on until power is restored.
When the power is restored, the control unit will return
to the previous comfort adjust setting in
Static
mode.
The
Turn
mode must be reselected manually.
Low Pressure Alarm
If the control unit senses a low pressure condition,
such as a disconnected hose or therapy cell, then it
will activate an audible alarm and will flash
“L”
and
“P”
on the display. Once the low pressure condition has
been corrected, the alarm will cease and the control
unit will return to normal operation.
Optional Accessories
• Additional therapy pads – available for purchase
Troubleshooting
Therapy Surface is Not Inflating
•
Ensure the hose connection from the therapy
mattress system (mattress) to the control unit is
securely connected.
• Ensure that the control unit is plugged into an AC
outlet.
•
Ensure that the power is not on
Standby
. If on
Standby
, press the
Power
button.
•
Ensure that all air cells are connected to the
internal mattress manifold.
Unable to Change Therapy Mode or Adjust
Comfort Control
Make sure the
Lockout
function is disabled. To
disable, press the up and down
Comfort Adjust
arrows
simultaneously.
Troubleshooting Modes of Operation
The DermaFloat
®
LR is designed to offer continuous
bilateral or unilateral rotation.
Continuous bilateral rotation (Turning the resident
to both left and right)
1.
Ensure the mattress is fully inflated with the
resident held centrally on the support surface
and correctly supported. If necessary, make
adjustments using the comfort control buttons.
2. Check to make sure the control panel is not locked
out. If the
Power
LED is flashing, then press the
Comfort Control
arrows together to unlock the
system.
3. Verify the system is in
Static
mode (
Turn
mode
None
LED will be lit).
4. Press the
Turn
mode button until the
Both
LED is
lit.
5. Press the
Turn Angle
button until the desired target
angle LED is illuminated.
6. Select the
Turn Cycle
time by pressing the
Turn
Time
button until the desired time period is
illuminated.
9