Joerns Support Surface

DermaFloat LR Model

© 2015 Joerns Healthcare • 6110064 RevB • 15-2003

Grounding Instructions

Warning: 

Use a properly grounded, three-prong, 120V 

AC outlet for this product. Failure to use a grounded 

outlet could result in personal injury or damage to 

equipment or house wiring, including risk of fire. A 

qualified electrician should be contacted to correct the 

wiring and ensure a properly  

grounded outlet.

Before installing this product, have the electrical 

system checked to make sure the electrical circuits 

and the electrical service are properly grounded.

Having a three-prong outlet does not necessarily 

mean it is grounded. Sometimes two-prong outlets are 

replaced with a three-prong type even though there is 

no ground wire. 

There is always a chance of a loose connection or 

poor installation of a ground wire that causes the loss 

of proper ground at the outlet. Inadequate grounding 

at electrical outlets can occur even if there is a 

ground wire. Wires can become loose over time at the 

connection to the outlet.

Note:

 To install new wires on a circuit requires a 

qualified electrician. 

How to Determine if Your Outlet has the 
Proper Grounding

Most hardware stores sell circuit testers (Figure 2) that 

can be used to test an outlet for proper grounding. 

The tester plugs into an outlet and by observing 

the indicator lights you can determine if the outlet is 

properly grounded. For a higher level of assurance, an 

electrician should be requested to thoroughly test the 

electrical system with more reliable equipment. 

If repair or replacement of the cord or plug is 

necessary, please contact Joerns Healthcare for 

assistance.

7

Setup

Warning:

 For important precautions, please see 

page two.

Caution:

 Do not place the 

control unit on the floor. 

Position the power cord to 

keep personnel from tripping 

over it. 

Note:

 The DermaFloat LR 

system must be installed 

on bed frames that are 

equipped with side rails. 

Please raise side rails on 

the bed and lock them in 

position after the resident 

is on the mattress. Never 

leave resident unattended 

on mattress system with bed 

side rails in the  

down position.

When the product is not in use, properly store 

the power cord. Failure to do so could result in 

personal injury.

• 

Remove the existing mattress from the bed.

•  Place the DermaFloat LR mattress with the 

hose connection at the foot end of the bed and 

the therapy cells facing up. Secure the straps 

on the mattress securely to the movable part 

of the bed frame.

• 

If the therapy pad is not already on the 

mattress, place it on the mattress. Attach the 

elastic straps to the mattress buckles around 

each corner of the mattress. Attach the six (6) 

additional straps to the movable part of the 

bed frame.

• 

Hang the control unit on the foot of the bed 

facing away from the bed. Attach the hose 

connector marked 

CPR 

to the control unit.

• 

Plug in the control unit and the press the 

Power 

key. The control unit will start and the 

Power

 light will illuminate. 

•  Allow up to a minute for full inflation. Place 

the resident on the mattress. Mattress can be 

inflated with resident on it, but will take longer, 

depending on their weight/size. 

Note: 

Keep 

the control unit on while the resident is on  

the mattress.

Figure 2

Содержание DermaFloat LR

Страница 1: ...User Service Manual Joerns Support Surface DermaFloat LR Model To avoid injury read user s manual before using...

Страница 2: ...presence of smoking materials or open flame Air flowing through the air mattress will support combustion 2 Danger To reduce the risk of shock adhere to the following instructions Failure to do so cou...

Страница 3: ...manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Jo...

Страница 4: ...ort Power Failure Low Pressure Alarm Optional Accessories Troubleshooting 9 10 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Troubleshooting Modes of Operation...

Страница 5: ...with the attending physician 5 Introduction The DermaFloat LR provided by Joerns Healthcare is a unique continuous bilateral rotation mattress replacement system This system is designed to assist in t...

Страница 6: ...of cleaning re assembly and cost effective service Lower two inch therapy support cell provides additional therapeutic support and remains inflated for up to 12 hours in the event of a power failure...

Страница 7: ...cal system with more reliable equipment If repair or replacement of the cord or plug is necessary please contact Joerns Healthcare for assistance 7 Setup Warning For important precautions please see p...

Страница 8: ...prior to Autofirm either Turn or Static in approximately 15 minutes Pause Use this feature to immediately stop all rotation and hold the current mattress position The mattress will hold this position...

Страница 9: ...e an audible alarm and will flash L and P on the display Once the low pressure condition has been corrected the alarm will cease and the control unit will return to normal operation Optional Accessori...

Страница 10: ...nent of urine and or stool and residents with heavily draining wounds Add top sheet blanket and or bedspread as needed for resident comfort Keep the amount of padding between the resident and bed to a...

Страница 11: ...age with possible body fluids it is recommend that the mattress system be cleaned with a mild detergent and warm water If disinfection is desired a combination cleanser disinfectant may be used as exp...

Страница 12: ...ithout Joerns s authorization will invalidate any warranties on this product Storage and Care Note Clean the DermaFloat LR as described in the previous section prior to storage Control Unit The power...

Страница 13: ...sive usage Improper maintenance handling and or cleaning Failure to use in the manner indicated in the DermaFloat LR user manual Any modification repair or alteration done to the DermaFloat LR that wa...

Отзывы: