![Joerns DermaFloat LR Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/joerns/dermafloat-lr/dermafloat-lr_user-and-service-manual_2021306008.webp)
8
Joerns Support Surface
DermaFloat LR Model
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
•
Position the resident’s head in the position that
the resident will be in for the largest portion of the
day. If the resident is lying flat, please use three
(3) fingers for the hand check. If the resident will
be sitting up for the majority of the day, please use
two (2) fingers.
• Perform a hand check by placing fingers locally
under the resident’s buttocks between two
cushions. The resident should not bottom out. If
they do, increase the therapy control by one LED
(light), until they no longer
bottom out.
• The safety side bolsters cannot be inflated unless
the
CPR
plug is installed. The plug is located
on the hose assembly between the mattress
and the control unit.
Caution:
The safety side
bolsters should always be inflated when resident is
unattended on the mattress and/or the mattress is
in
Turn
mode.
Operation
Warning:
For important precautions, please see
page two.
Caution:
The resident’s head should be positioned
in the center of the top section of the mattress. When
using the mattress system always ensure that the
resident is positioned properly within the confines of
the bed. Do not let any extremities protrude over the
side or between the bed rails when the mattress is
being used.
Resident Comfort Controls and Monitoring
Power
The
Power
button is used to turn the power on and off.
Autofirm
Autofirm
mode provides maximum air inflation
designed to assist both residents and caregivers
during resident transfer and treatment. The unit will
automatically return to the mode it was in prior to
Autofirm
(either
Turn
or
Static
) in approximately
15 minutes.
Pause
Use this feature to immediately stop all rotation and
hold the current mattress position. The mattress will
hold this position until this feature is turned off by
pressing the
Pause
button a second time.
8
Lockout
This unit is designed to lock out all of the adjustment
controls after the resident has been positioned
correctly. Ten minutes after the last button has been
pushed, push the
Power
light begins to flash indicating
Lockout
is enabled. This feature is to prevent any
unauthorized changes to the resident settings. To
unlock and make adjustments to the settings, press
both up and down comfort arrows at the same time.
Lockout mode will return after ten minutes.
Note:
The unit is designed to lock out all the
adjustment controls after the resident has been
positioned correctly. In approximately ten minutes after
the last button push the power on light begins to flash
indicating
Lockout
is enabled.
Comfort Adjust
The
Comfort Adjust
function is located across the
top of the control panel. The DermaFloat
®
LR can be
customized to meet individual resident needs within
a therapeutic window. Use the soft and firm keys to
simultaneously increase or decrease pressure in the
entire mattress in
Turn
and
Static
modes. The current
setting is shown in the LED display in the center of the
control panel.
Turn Mode
The
Turn
mode and sets the desired rotation type. The
resident can be rotated to either side individually or to
both sides in a continuous cycle.
Turn Time
Use the
Turn Time
feature to set the time cycle for
the selected
Turn
mode. The DermaFloat LR can be
placed into a 10, 20, or 60 minute cycle based on what
is determined to be the optimal therapy setting for the
individual resident.
Figure 3