 APM

Lockout

This feature is to prevent any unauthorized changes to 

the resident settings. To unlock and make adjustments 

to the settings press both up and down comfort arrows 

at the same time to disengage the 

Lockout

 function. 

The 

Lockout

 function will return in approximately three 

minutes.

Note: 

The unit is designed to lock out all the 

adjustment controls after the resident has been 

positioned correctly. In approximately three minutes 

after the last button push, the 

Power 

on light begins to 

flash indicating 

Lockout

 is enabled.

Additional Features

Warning:

 A possible fire hazard exists. This product is 

suitable for use with oxygen administering equipment 

of the nasal, mask, or half bed-length, tent-type only. 

To prevent personal injury or equipment damage, 

ensure that the oxygen tent does not extend below the 

mattress.

CPR

Disconnect the hose from the control unit. Deflation 

times will vary based on resident weight and profile. To 

resume therapy, reattach the hose to the control unit 

(Figure 3).

CairLock

The mattress system comes equipped with our 

specially designed CairLock therapy connectors to 

provide greater reliability and comfort. In the event 

of a power failure the connectors can be interlocked 

together to ensure full-continued therapy for 12 hours. 

Disconnect the two hoses from the control unit and 

interlock the two ends by pressing them together 

(Figures 4 and 5).

Transport

To transport the resident in bed, set mode to 

Therapy

 

and turn the control unit off. Unplug the power cord 

from the outlet. Transport can be done with the unit 

plugged in to the mattress or by using the CairLock 

feature.

Power Failure

In the event of a power failure, if power is restored 

within approximately one hour, the control unit will 

return to previous settings. If the duration of a power 

failure extends beyond an hour, the control unit will 

default to 

Standby

 mode when power is restored. 

Reset the preferred mode and comfort level as 

described in the “Operation” section. In the event of a 

Figure 4

Figure 5

Figure 3

Содержание DermaFloat APM

Страница 1: ...Joerns Healthcare User Service Manual Joerns Support Surface DermaFloat APM transforming wound management To avoid injury read user s manual before using...

Страница 2: ...presence of flammable anesthetics Do not use in the presence of smoking materials or open flame Air flowing through the air mattress will support combustion Danger To reduce the risk of shock adhere...

Страница 3: ...esponsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and produc...

Страница 4: ...when equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the head panel foot panel and bed or side rails The equipment is NOT to be used when such gaps are...

Страница 5: ...ts should be considered on an individual basis and discussed with the attending physician Features The DermaFloat APM is comprised of two components Therapy control unit Therapy mattress system Therap...

Страница 6: ...though there is no ground wire There is always a chance of a loose connection or poor installation of a ground wire that causes the loss of proper ground at the outlet Inadequate grounding at electric...

Страница 7: ...de Operation Warning For important precautions see page two Caution The resident s head should be positioned in the center of the top section of the mattress When using the mattress system always ensu...

Страница 8: ...ll vary based on resident weight and profile To resume therapy reattach the hose to the control unit Figure 3 CairLock The mattress system comes equipped with our specially designed CairLock therapy c...

Страница 9: ...or the highest comfort control setting to achieve quick inflation Locate the Ready Valve by unsnapping the flap on the hose set Turn the Ready Valve to inflate or deflate as desired Return the control...

Страница 10: ...s of resident migration or inadvertent deflation of resident surface Skin Care Monitor skin conditions regularly particularly in areas where incontinence and drainage occur or collect and consider adj...

Страница 11: ...hands Therapy Pad The therapy pad can be wiped down with a disinfectant solution or a mild detergent with a damp cloth If heavily soiled the therapy pad can be laundered in a washer and dryer with wa...

Страница 12: ...ght must not exceed bed frame manufacturers specified load capacity Environmental Conditions Operating Conditions Ambient temperature 10 C to 40 C Relative humidity Non condensing Storage And Shipping...

Страница 13: ...13 Joerns Support Surfaces DermaFloat APM Notes...

Страница 14: ...essive usage Improper maintenance handling and or cleaning Failure to use in the manner indicated in the DermaFloat APM user manual Any modification repair or alteration done to the DermaFloat APM tha...

Отзывы: