background image

7

Joerns

®

 Support Surfaces

DermaFloat

®

 APM

•  Select the appropriate 

Comfort Adjust

 level to 

prevent bottoming out (i.e., providing greater than 

one inch of air between the resident’s sacral area/

buttocks and the lower safety mattress) as outlined 

below:

1.  Begin by placing the head of the bed in the 

appropriate position based on the resident’s 

clinical condition. 

2.  Select the highest or most firm 

Comfort Adjust

 

setting.

3.  Hand Check: Place a hand with three (3) fingers 

(if head of bed at 30° or higher) or four (4) 

fingers (if head of bed lower than 30°) stacked 

vertically beneath the cells of the mattress and 

above the safety mattress directly between 

the lowest point of the resident’s sacral area/

buttocks. The smallest finger should be resting 

on the safety mattress. 

4.  Sequentially reduce the 

Comfort Adjust

 setting 

to the firmness level where the height of the 

three (3) or four (4) fingers can slide with 

minimal resistance between the resident’s sacral 

area/buttocks and the lower safety mattress. 

This is the proper 

Comfort Adjust

 setting for the 

resident to assure proper inflation of the air cells 

and prevent bottoming out of the mattress.

5.  Document the resident’s 

Comfort Adjust

 setting 

for future reference, and re-evaluate with 

the Hand Check as the resident’s condition 

warrants.

•  The CairRails risk management side air bolsters 

can be inflated or deflated as required. Locate 

the turn valve on the hose assembly between the 

mattress and the control unit. Next, inflate/deflate the 

CairRails by moving the turn valve to the up (inflate) 

or down (deflate) position. 

Note:

 When inflating or deflating CairRails it is 

recommended that the support surface be in 

Autofirm

 

mode.

Operation

Warning:

 For important precautions, see page two.

Caution:

 The resident’s head should be positioned 

in the center of the top section of the mattress. When 

using the mattress system always ensure that the 

resident is positioned properly within the confines of 

the bed. Do not let any extremities protrude over the 

side or between the bed rails when the mattress is 

being used. 

Resident Comfort Controls and 
Monitoring (Figure 2)

 

Power

The 

Power

 button is used to turn the power on and off.

Standby

The unit starts up in 

Standby

. Press the 

Power 

button 

to inflate the mattress. When the 

Standby 

light is 

on, it may also indicate that there has been a power 

interruption and the therapy control unit is ready to be 

turned back on. Press the 

Power

 button and reset the 

preferred mode of therapy and comfort level.

Modes

Autofirm

Autofirm

 mode provides maximum air inflation 

designed to assist both residents and caregivers 

during resident transfer and treatment. The unit will 

automatically return to the mode it was in prior to 

Autofirm

 (either 

Therapy

 or 

Alternate

) in approximately 

10 minutes.

Therapy Modes: Therapy and Alternating 

•  Therapy

: The unit starts in the 

Therapy

 mode, which 

is the standard low air loss therapy. 

•  Alternate

: The air cylinders are inflated and deflated 

in an A-B-A-B pattern; that ensures the resident is 

always supported.

Cycle Times

The 

Cycle Time

 function is located on the bottom 

right of the control panel. Select the frequency of the 

alternation of the therapy cells in 5, 10 or 15 minutes 

cycles. For more aggressive therapy, choose the five 

minute option. 

Comfort Adjust

The 

Comfort Adjust

 function is located in the center 

of the control panel. The DermaFloat

®

 APM can be 

customized to meet individual resident needs within 

a therapeutic window. See “Setup” section for more 

information. This function will not work in 

Autofirm

 

mode. 

Figure 2

Содержание DermaFloat APM

Страница 1: ...Joerns Healthcare User Service Manual Joerns Support Surface DermaFloat APM transforming wound management To avoid injury read user s manual before using...

Страница 2: ...presence of flammable anesthetics Do not use in the presence of smoking materials or open flame Air flowing through the air mattress will support combustion Danger To reduce the risk of shock adhere...

Страница 3: ...esponsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and produc...

Страница 4: ...when equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the head panel foot panel and bed or side rails The equipment is NOT to be used when such gaps are...

Страница 5: ...ts should be considered on an individual basis and discussed with the attending physician Features The DermaFloat APM is comprised of two components Therapy control unit Therapy mattress system Therap...

Страница 6: ...though there is no ground wire There is always a chance of a loose connection or poor installation of a ground wire that causes the loss of proper ground at the outlet Inadequate grounding at electric...

Страница 7: ...de Operation Warning For important precautions see page two Caution The resident s head should be positioned in the center of the top section of the mattress When using the mattress system always ensu...

Страница 8: ...ll vary based on resident weight and profile To resume therapy reattach the hose to the control unit Figure 3 CairLock The mattress system comes equipped with our specially designed CairLock therapy c...

Страница 9: ...or the highest comfort control setting to achieve quick inflation Locate the Ready Valve by unsnapping the flap on the hose set Turn the Ready Valve to inflate or deflate as desired Return the control...

Страница 10: ...s of resident migration or inadvertent deflation of resident surface Skin Care Monitor skin conditions regularly particularly in areas where incontinence and drainage occur or collect and consider adj...

Страница 11: ...hands Therapy Pad The therapy pad can be wiped down with a disinfectant solution or a mild detergent with a damp cloth If heavily soiled the therapy pad can be laundered in a washer and dryer with wa...

Страница 12: ...ght must not exceed bed frame manufacturers specified load capacity Environmental Conditions Operating Conditions Ambient temperature 10 C to 40 C Relative humidity Non condensing Storage And Shipping...

Страница 13: ...13 Joerns Support Surfaces DermaFloat APM Notes...

Страница 14: ...essive usage Improper maintenance handling and or cleaning Failure to use in the manner indicated in the DermaFloat APM user manual Any modification repair or alteration done to the DermaFloat APM tha...

Отзывы: