2   Safety Instructions 

Flow switch 

2   Safety Instructions 

Always read these operating instructions carefully prior to installing the new product. Always adhere to the 
instructions contained herein, especially the safety instructions, otherwise there is a potential risk of personal 
injury and damage to instruments and plants. 

Even though 

JLSO 

provides assistance through personal consultation or the respective literature, it is the 

responsibility of the customers to determine the suitability of the product for the specific application. 
The flow switches are state-of-the-art devices. This concerns switching point accuracy, functioning and safe 
operation of the device.

 

However, professional and safety conscious conduct of the operator is required to ensure safe operation. 

2.1   Qualified personnel 

The personnel entrusted with installing, operating and maintaining the flow switches have to be suitably
qualified; the required knowledge can be gained via training courses or appropriate on-the-job instruction.
The personnel have to be familiar with the contents of these instructions, which have to be available to
them at all times.

The electrical connection should only be carried out by a fully qualified electrician.

All work has to be carried out in accordance with existing national regulations on accident prevention and
safety at work and with any internal regulations of the operator, even if they are not specified in these
instructions.

2.2   Special safety instructions 

The operational safety of the supplied equipment is only guaranteed if it is operated according to its
intended use. The specified limit values (see the chapter entitled “Technical Data“) should never be
exceeded.

To avoid damages to the flow switch and the monitored system, only use 

JLSO

 flow switches for minimum

or maximum monitoring of the flow of liquids.

Always follow and adhere to the flow switch installation  instructions.

Never operate the flow switch in systems which have a greater flow rate than the specified max. flow rate
(see the chapter entitled “Technical Data”, details of the max. flow rate). Otherwise it will cause
irreparable damage to the flow switch.

Prior to flow switch installation, ensure that all the materials of the flow switch are chemically and
mechanically-resistant to the medium which is to be monitored and to all external factors.

Ensure that the medium is free from magnetic particles.

Suitable measures should be taken to prevent the medium from freezing.
If the flow switch is to be used in ambient temperatures of <4 °C, do not carry out any operation
beforehand with pure water, e.g. a test run. Residual water in the flow switch can result in frost damage.

No greases, oils etc. should be used during the installation of the VK... devices due to the material
resistance.

Ensure that the max. specified operating pressure is not exceeded.

Never remove a flow switch or its upper parts from a pipe system under pressure.

If the medium which is to be monitored is very hot, the flow switches or their connection fittings will also
become very hot. In this case, neither touch the flow switch nor place any heat-sensitive objects in its
vicinity.

Protect the flow switch against external magnetic fields in the immediate vicinity, since these can impair
device functioning.

The technical data of special versions (customised versions) can deviate from the details in these
instructions. Please observe the details on the type plate.

Page 4 of 15

Copyright ©  Jlso Tec Trade GmbH

JLSO Tec Trade GmbH 

In den Birken 98

Fon:

 

+49 (0) 

6369 230 9961

 

[email protected]

Fax:

 

+49 (0) 

6369 230 9963

 

https://jlso-tec-trade.de

Daniel

 

H.

 

Meckel

D-66999

 

Hinterweidenthal

Содержание VH M 1 1 1 Series

Страница 1: ...age 1 15 www jlso tec trade de Page 1 of 15 Copyright Jlso Tec Trade GmbH JLSO Tec Trade GmbH In den Birken 98 Fon 49 0 6369 230 9961 info jlso tec trade de Fax 49 0 6369 230 9963 https jlso tec trade...

Страница 2: ...g connector EN 175301 803 A 25 4 2 3 Sensor plug M12x1 4 pole 26 4 2 4 Fixed connecting cable 26 5 Adjusting the Switching Unit 27 5 1 Contact type 27 5 1 1 Standard contact 27 5 1 2 Change over conta...

Страница 3: ...sion VH X is designed for application in explosive atmospheres The ignition energy of the explosive atmosphere should not be below 60 J Please observe the following separation of zones The constructio...

Страница 4: ...witch and the monitored system only use JLSO flow switches for minimum or maximum monitoring of the flow of liquids Always follow and adhere to the flow switch installation instructions Never operate...

Страница 5: ...ive loads Details of the protective circuit can be requested from the manufacturer 3 Material Specifications of Wetted Components Type VH M 1 1 1 VH M 1 1 C VH M 1 1 1X VH M 3 3 3 VH M 3 3 3X VH M P 1...

Страница 6: ...re desired Please make sure that there are no external magnetic fields in the immediate vicinity of the flow switch since these can impair device functioning Fig 2 There is an arrow on the flow switch...

Страница 7: ...tallation with the help of brass soldering threaded nipples or stainless steel welding threaded nipples including O ring V 05M V 06M up to DN 50 V 07M up to DN 50 Fig 4 V 06M from DN 50 V 07M from DN...

Страница 8: ...before connecting the wires of the mains cable CAUTION Ensure that the maximum electrical contact load specified on the type plate is never exceeded otherwise the reed contact integrated in the switc...

Страница 9: ...unction box pos 2 Tighten the cable gland PG 9 pos 5 Place the junction box pos 2 on the connector pos 1 and retighten the central screw pos 6 Fig 7 Fig 8 Fig 9 Fig 10 To guarantee the protection clas...

Страница 10: ...ing to the connection diagram see Fig 13 or Fig 14 Standard contact Flow switch version VH3 VK3 Flow switch version VH3 X Fig 13 To prevent electrostatic charging the VH3 X devices have to be connecte...

Страница 11: ...wise agreed with the customer the switching unit is factory set as a make contact i e the reed contact opens if the set switching point is exceeded 5 1 2 Change over contact only for VH3 For flow swit...

Страница 12: ...ried out on the basis of the arrow length Movement towards the arrow head Switching point is set to lower flow rate Movement towards the arrow tail Switching point is set to higher flow rate Carefully...

Страница 13: ...wise agreed with the customer the switching unit is factory set as a make contact i e the reed contact opens if the set switching point is exceeded 5 1 2 Change over contact only for VH3 For flow swit...

Страница 14: ...deviate from the details in these instructions Please observe the details on the type plate Reduced pressure level for devices with copper pipe section Please observe the details on the type plate 8...

Страница 15: ...rate m h DN 50 30 25 DN 80 80 65 DN 100 150 100 V 05M mounting length 51 mm DN 150 200 170 DN 100 100 40 DN 150 150 95 V 06M mounting length 111 mm DN 200 200 160 DN 50 30 DN 80 100 DN 100 150 V 01M...

Отзывы: