Functions
Description
Register
Modbus
reference
Address
Product address (R/W)
0x0081
40130 (0x0081)
Serial number
Serial number of the product
0x0030~0x0035
40049 (0x0030)
Flow rate
Current flow rate (R)
0x003A~0x003B
40059 (0x003A)
Accumulated flow
Accumulated or totalized flow rate (R)
0x003C~0x003E
40061 (0x003C)
Baud rate
Communication (R/W)
0x0082
40131 (0x0082)
GCF*
Gas correction factor (R/W)
0x008B
40140 (0x008B)
Response time*
Set the response time (R/W)
0x008D
40142 (0x008D)
Units*
Instant and accumulated flow rate units (R/W) 0x0090
40145 (0x0090)
Password*
Password change (R/W)
0x00AE~oxooAF 40175 (0x00AE)
Reset offset*
Reset or calibrate the offset (W)
0x00F0
40241 (0x00F0)
Reset totalizer*
Reset accumulated or totalized flow rate (W)
0x00F2
40243 (0x00F2)
Write protection
Write protection of selected parameters (W)
0x00FF
40256 (0x00FF)
Notes:
1, R
–
Read-only, W
–
Write only, R/W
–
Read and write.
2, For the * marked functions, need to disable the write protection before setting.
The detailed information of each register is described below: Y: enabled; N: disabled
Address
0x0081
Write
Y
Read
Y
Description
RS485 Modbus address of the flow meter
Value type
UINT 16
Notes
Values from 1 to 255 except for 157 (0x9d).
Broadcast address 0 is not enabled, the default address is 1.
SN, Serial number
0x0030
Write
N
Read
Y
Description
Series Number of the product, SN
Value type
UINT8 (12 bits)
Notes
SN=
value(0x0007), value(0x0008),….,value (0x000C);
e.g.:
Receiving 12 bits as: 0x2A47, 0x3741, 0x4549, 0x3032, 0x3035, 0x382A ,
the corresponding Serial Number is
*G7AEI02058*
.
Instant flow rate
0x003A ~ 0x003B
Write
N
Read
Y
Description
Instant flow rate
Value type
UINT 16
Notes
Flow rate = [Value (0x003A)*65536 + value (ox003B)]/1000
e.g.:
For a flow rate of 20.340 L/min
, the user will read “0
(0x0000)
” from
register 0x003A and “20340
(0x4F74)
” from register 0x003B,
therefore
Current flow rate = (0*65536+20340)/1000 = 20.340
25
|
P
a
g
e
http://www.jetec.com w
om.tw/
.t
電話:
0
4
-
23729418 傳真:04-23724011 Email:[email protected] LINE:@jetec
MF5000 User Manual