Device Installation and Connection | Operation & Maintenance Instructions TOPAX L1/L4/L5 |
13
.4 terminal connection diagram of built-in casing
Electrode/
Clips
Potentiostatic
Type PM-J
Potentiostatic-
Type PM-K
Amperometric
Hydrogen
peroxide
1 = Shield
2 = Measuring-
3 = Reference-
4 = Counter-
electrode
= -
= Gold- (violet)
= (black)
= stainless steel- (red)
elektrode
= -
= (Pin 1)
= (Pin 2)
= (Pin 3)
= -
= Platin- (rot)
= Brigde to 4
= Copper- (blue)
or Silver electrode
= Shield
= (brown)
= (white)
= green (blue)
TOPAX L1 (Cl
2
)
X (<4 mg/l)
X (>4 mg/l)
X
-
TOPAX L4 (ClO
2
)
X (<4 mg/l)
X (>4 mg/l)
-
-
TOPAX L5 (H
2
O
2
)
-
-
-
X
Connection
Terminals
Directions
Pt100
5 + 6
Analog Output
9 + 10
9 = +, 10 = -, maximum 500 Ohm load
Digital Input
11 + 12
11 = +, 12 = -, external controller stop or water deficiency
Relay 1
18 + 19
Relay 2
20 + 21
Relay 3
22 + 23
Alarm-Relay
Power Supply
24 + 25 + 26
Check indications on rating plate!
RS485
(Optional)
Sub-D
3 = +, 8 = -
Sub-D 4/7
Bridge circuits = termination impedance activated
Terminal connection of electrodes
Potentiostatic electrode, Type PM-J, <4 mg/l
1
2
3
5
4
6
7
8
9
10
11
12
13
14
1
2
3
4
5
9
8
7
6
15
16
17
18
19
20
21
22
23
L
N
PE
Relay 1
Relay 2
Relay 3
24
25
26
Measurement, Gold electrode (violet)
Reference electrode (black)
Counter
, stainless steel electrode (red)
RS 485
(OPTION)
Measurement, platinum electrode (red)
Copper (blue) or silver electrode (violet)
1
2
3
5
4
6
7
8
9
10
11
12
13
14
1
2
3
4
5
9
8
7
6
15
16
17
18
19
20
21
22
23
L
N
PE
Relay 1
Relay 2
Relay 3
24
25
26
RS 485
(OPTION)
Shield
1
2
3
5
4
6
7
8
9
10
11
12
13
14
1
2
3
4
5
9
8
7
6
15
16
17
18
19
20
21
22
23
L
N
PE
Relay 1
Relay 2
Relay 3
24
25
26
RS 485
(OPTION)
Measurement electrode (brown)
Reference electrode (white)
Counter electrode (green / blue)