6 MEASUREMENT OF RELAXATION TIME
Arrow buttons
Fig. 6.1
Set interval window
3.
Click the
Set Value
button.
The array parameter window closes, and the values you set are entered into
tau_interval
in the
Experiment Tool
.
4.
Click the
Submit
button.
The measurement starts.
Although the data-processing procedure is the same as that for the inversion-recovery
method, use the following approximation formulas.
Weighted Linear Sat. Recovery: The weighted linear least-squares method
The weight is reduced for the measurement having
long
τ
.
Unweighted Linear Sat. Recovery: Linear least-squares method without weights.
Nonlinear Saturation Recovery: Nonlinear
least-squares
method
Refer to the user’s manual “Application” for information about the measurement and
processing of the inversion-recovery method.
NMECAXS_V50-SLD-2