3 SETTING OF INSTRUMENT CONDITIONS
7.
Perform data processing.
Refer to the user’s manual “Data processing” for processing the data obtained
by the array measurement.
After data processing, the result is displayed by selecting
Linearize
.
8.
Set up the
obs_amp_grad
value of
Carbon13
.
In the spectra for which you carried out
Linearize
processing, the value of
obs_amp_grad
with the strongest signal intensity is the optimum value of cross-po-
larization conditions.
9.
Perform array measurement of
Carbon13
to set
obs_amp_cp
.
3.7.4 Selecting a Method for Cross Polarization
There are several methods for the cross polarization in the CPMAS measurement. You
can change the method by choosing combinations in
irr_shape_cp
and
obs_shape_cp
.
irr_shape_cp
obs_shape_cp
Constant CP
constant_cp
constant_cp
constant_cp RAMP_cp
RAMP CP
RAMP_cp constant_cp
constant_cp APHH_cp
APHH CP
(Adiabatic Passage Hartmann Hahn CP)
APHH_cp constant_cp
ad freq sweep CP
(ADiabatic FREQuency sweep CP)
ad_freq_cp constant_cp
constant_cp fm_cp
fm_cp constant_cp
FM CP
(Frequency Modulation CP)
fm_cp fm_cp
nicp
NICP
(Nuclear Integrated CP)
sadis_cp
constant_cp
nicp
sadis CP
(Simultaneous ADIabatic frequency
Sweep)
sadis_cp
sadis_cp
3-12