4 ADJUSTMENT OF MEASUREMENT CONDITIONS AND PROCESSING CONDITIONS
4.5 SETTING SAMPLING POINT (X_POINTS)
The parameter
x_points
specifies the number of sampling points to acquire FID data.
For a solid-state NMR measurement, this parameter is used to adjust the decoupling time.
Since decoupling time is the same as acquisition time, it is expressed as
x_acq_time
.
High-power RF will be input to the probe during
x_acq_time
.
In order to reduce the load of the probe, it is necessary to shorten
x_acq_time
. However,
if
x_acq_time
is too short, the FID signal will not be acquired. Ideally it is best that
x_acq_time
is as short as possible while still acquiring the entire FID signal. A good
example and a bad example are shown in the following figure. As estimates,
x_acq_time
is about 20 ms in polymers and about 40 to 100 ms in crystals.
Good example
Undesirable example:
x_acq_time
is too long.
Bad example:
x_acq_time
is too shor t
Fig. 4.4 Good and bad examples of acquisition time
NMECAXS_V50-SLD-2