3 SETTING OF INSTRUMENT CONDITIONS
9.
Click the
Sumit
button.
The pulse occurs, and the measurement starts.
When the measurement is completed, the nD Processor window automatically
opens.
10.
Perform data processing.
Refer to the user’s manual “Data processing” for instruction on processing the
data obtained by the array measurement.
After data processing, the result is displayed by selecting the
Linearize
.
11.
Find the 360
°
pulse width.
In the spectrum after the
Linearize
processing, the pulse width where the signal in-
tensity is nearest to 0 (zero) is the 360
°
pulse width. Furthermore, for longer pulse
widths, the peaks should be positive.
Fig. 3.4 Spectrum after carrying out Linearize processing
12.
Obtain the 90
°
pulse width as one fourth of the 360
°
pulse width you found.
■
Measuring the pulse width of the irradiation channel
For the same frequency, the spectrometer is adjusted so that the pulse width does not dif-
fer in the observation channel and the irradiation channel. Therefore, it is not necessary to
measure the pulse width of the irradiation channel.