Err 5
Fatal
Light Saturation Not Found
This error indicates th
at the peak light hasn’t been found at
zero.
The most likely causes of this error are:
1. Lamp failure.
2. Deteriorating lamp signal.
3. Sample or cuvette in the sample holder.
Solution:
Ensure the sample holder is empty. Restart the unit, if the
problem persists contact a service technician.
Err 8
Warning
Redo
Unable to find Vane on Turret
This error indicates that the vane position zero on the turret
cannot be found.
The most likely causes for this error are:
1. The turret carousel has been removed and not replaced.
Solution:
Check if the turret is in the sample chamber and inserted
properly. Press the key adjacent to the back icon to try
rechecking.
Err 9
Fatal
Over Temperature
This error indicates that the thermal switch has cut out.
The most likely causes for this error are:
1. Failure of the fan
2. Thermal switch not connected.
Solution:
Restart the unit, if the problem persists contact a service
technician.