42
For sipping
:
1. Connect the sipper pump tubing to the outlet
port on the flow-through cuvette.
2. Secure the tubing using the clip on the
righthand side of the pump head.
3. Ease the tubing round the rollers by carefully
rotating them clockwise, by hand. Clamp the
tubing into the clip on the left hand side of the
motor.
4. Once secured, ensure the tubing is routed into the two retaining clips located on the base plate at
the side of the pump head.
5. Cut the tubing at the point where it fits comfortably onto the left hand tube located on the inside of
the front bulk head.
6. Connect a suitable length of this tubing to the external waste pipe.
7. Cut a small length of the sipper pump tube and push this over one end of the capillary tube.
Connect this to the inlet port of the flow-through cuvette.
8. Route the tube into the two retaining clips located on the base plate at the side of the pump head.
9. Fit the sipper probe and secure using the thumbscrew. Feed the capillary tubing through the tube
and up through the sipper probe, allowing sufficient length for it to pass into a suitable receptacle.
For pumping
:
1. Cut two pieces of sipper pump tubing
approximately 300mm in length. Take one
length of tubing and fit this to the pump head,
as shown, securing the tubing using the clip on
the right hand side of the pump head.
2. Ease the tubing round the rollers carefully
rotating them clockwise, by hand. Clamp the
tubing into the clip on the left hand side of the
motor.
3. Fit the other end onto the inlet port on the
flow-through cuvette.
4. Fit the second 300mm length of tubing to the outlet port of the flow-through cuvette. Once