Temperature measurement
52
By varying the offset value, the user can adjust the temperature display
of the camera to the actual temperature value. As this offset acts on the
radiant power on which temperature calculation is based, unknown
fractions of extraneous radiation are easily compensated.
object
camera
1
2
3
environment
air path
(1) Direct radiation of measured object:
(
)
λ
∗
τ
∗
ε
,
T
P
o
P
o
with
1000
ce
tan
dis
ap
p
e
∗
−
=
τ
(2) Ambient radiation reflected by measured object:
(
)
(
)
λ
∗
τ
∗
ε
−
,
T
P
1
A
p
o
(3) Characteristic radiation of air path:
(
) ( )
λ
∗
τ
−
,
T
P
1
p
p
Total radiation:
(
) (
)
(
)
(
) ( )
u
p
p
A
p
o
o
p
o
1
O
,
T
P
1
,
T
P
1
,
T
P
P
+
λ
∗
τ
−
+
λ
∗
τ
∗
ε
−
+
λ
∗
τ
∗
ε
=
When you switch-on the VARIOSCAN
high resolution
, a calibration file
is loaded containing the temperature-relevant parameters of the current
camera configuration. This file also contains the thermal characteristic
of the camera set up of 20 calibration points distributed over the entire
temperature measuring range and the corresponding conditions of
calibration. Subsequently, all other temperatures are derived by
InfraTec
Содержание VARIOSCAN high resolution Series
Страница 1: ...VARIOSCAN high resolution Models 3021 3021 ST and 3022 Manual JENOPTIK Group I n f r a T e c...
Страница 10: ...Technical Description 10 I n f r a T e c...
Страница 12: ...Technical Description 12 I n f r a T e c...
Страница 14: ...Technical Description 14 Total weight of system in carrying case approx 11 kg I n f r a T e c...
Страница 24: ...Operation 24 To switch off the activated magnifier again double click any of the zoom keys I n f r a T e c...
Страница 55: ...Maintenance 55 I n f r a T e c...