Temperature measurement
50
9
Temperature measurement
9.1
General
VARIOSCAN
high resolution
is a measuring slow-scan thermographic
system scanning the object field in a raster using a single element
detector.
The detector signal is amplified, digitised with 16 bits, and visualised
with an 8 bit resolution (256 colours). Every colour of the displayed
thermogram represents a defined temperature. Between two image
scans, a reference radiator (chopper) is moved into the optical path for
a short time, and the detector signal clamped to a defined voltage
value. In this way, the detector signal is referenced to an absolute
temperature. The temperature of the chopper and of the detector chip
are measured by sensors. To expand the temperature range of the
camera, different apertures can be swung into the optical path in front
of the detector.
The relationship between detector signal and object temperature is
non-linear. It is described by a family of characteristics based on
chopper and detector temperature. This family of characteristics is
saved in a data record of camera-specific SPC files in the form of
backing points at different object, chopper and detector temperatures.
These files are the result of a calibration. During calibration, every
camera head is measured point-by-point over the full temperature
measuring range at different ambient temperatures and using the
individual apertures. From the calibration points, a special interpolation
method generates a solid characteristic. All temperature-relevant
calculations are based then on this characteristic.
Temperature calculation on VARIOSCAN
high resolution
is performed
in three steps:
1. The current characteristic of the camera head is calculated based
on calibration data (SPC file) and depending on current ambient
conditions (chopper and detector temperature). As instrument
temperature changes relatively slowly only, the characteristic is
recalculated only on system start and at defined time intervals
(refresh).
2. Next, based on this characteristic the relationship is set up between
pixel signal (digitised voltage value of the amplified detector signal)
InfraTec
Содержание VARIOSCAN high resolution Series
Страница 1: ...VARIOSCAN high resolution Models 3021 3021 ST and 3022 Manual JENOPTIK Group I n f r a T e c...
Страница 10: ...Technical Description 10 I n f r a T e c...
Страница 12: ...Technical Description 12 I n f r a T e c...
Страница 14: ...Technical Description 14 Total weight of system in carrying case approx 11 kg I n f r a T e c...
Страница 24: ...Operation 24 To switch off the activated magnifier again double click any of the zoom keys I n f r a T e c...
Страница 55: ...Maintenance 55 I n f r a T e c...