OPERATOR
Instructions
Pipetting
To aspirate
1 Carefully hold the pipette and insert it securely into the polypropylene nose cone and silicone holder.
2 Select mode of operation – High, Low, or ‘G’.
3 Depress upper trigger (1) (The harder the pipetting triggers are pressed the faster the
aspiration.)
To dispense
1 In BLOW-OUT mode, (mode switch on high or low) depress the lower trigger (2) (The
harder the pipetting triggers are pressed, the faster the dispense).
2 If ‘G’ gravity mode is chosen, the upper aspirate trigger automatically operates at low speed
while the lower trigger opens the pipette to atmosphere, allowing liquid to run out.
WARNING!
Please ensure that no liquid enters the instrument. The
POWERPE TTE
/
POWERPE TTE
Plus
/
POWERPE TTE
should never be used without the filter (8).
The filter also prevents liquid from entering the instrument. Where liquid has entered please
clean and replace filter.
Trouble Shooting
The
POWERPE TTE
/
POWERPE TTE
Plus
/
POWERPE TTE
are virtually
service-free. If the recommendations above do not resolve the problem, please contact your
authorized dealer.
Note
When the meniscus level has been reached liquid should not drip out of the pipette. If the
pipette does drip, check that the adaptor assembly and filter are correctly assembled and the
pipette is properly located into the silicone pipette holder.
TROUBLE
POSSIBLE CAUSE
REMEDY
Filter blocked, dirty or damaged
Clean or replace
Suction capacity slow
membrane filter
Batteries low
Recharge
Pipette not securely positioned
Re-attach
Pipette drips
in silicone pipette holder
Pipette holder or pipette damaged
Replace