8

B. Conductivity Measurements

1.

Turn the unit on. Place the probe in the solution to be

measured. Immerse the probe (at least 2” to 3” or 5~7cm from

the tip). Shake the probe lightly to remove any trapped air

bubbles in the conductivity cell.

2.

Press

“Mode”

key to enter the desired measurement mode

(Conductivity, TDS or Salinity). The message “over” or “undr”

may appear briefly on the display indicate auto-ranging; this is

normal. Allow temperature to stabilize before taking

measurements.

C. Preparing Standard Solutions

Suitable conductivity standards are available commercially or

the user can prepare them using research grade reagents.

Here are some standard solutions the user can prepare to

calibrate the probe of the model 3177.

1. Standard solution of 1413 uS/cm at 25

: Accurately weight

out 0.746 grams of research grade dried Potassium Chloride

(KCL). Dissolve in 1000 ml of distilled water.

2. Standard solution of 12.90 mS/cm at 25

: Accurately weight

out 7.4365 grams of research grade dried Potassium Chloride

(KCL). Dissolve in 1000 ml of distilled water.

3. Standard solution of 111.9 mS/cm at 25

: Accurately weight

out 74.264 grams of research grade dried Potassium Chloride

(KCL). Dissolve in 1000 ml of distilled water.

[Note:

You can store the unused portion of the standard

solution in a plastic container for six months but the air space

between the cap and the solution must be kept to an absolute

minimum. Storing the excess solution below 4

can increase

the storage life. If you have any doubt of the accuracy of the

stored solution, a fresh batch should be prepared.

]

Содержание VisionPlus 3177

Страница 1: ...0 Operation Manual MODEL 3177 Microcomputer Based Conductivity TDS Salinity Temperature Benchtop Meter 3177...

Страница 2: ...OWER INPUT 2 INSTALLING THE BATTERIES 3 DISPLAY KEYS FUNCTIONS 4 A Display 4 B Keys 5 OPERATIONAL PROCEDURES 6 A Calibration 6 B Conductivity Measurements 8 C Preparing Standard Solutions 8 ERROR DISP...

Страница 3: ...just all the parameters for conductivity and will also give the user the option to select four types 0 01 0 1 1 0 10 0 of probe cell constant for a better selection of available probes and application...

Страница 4: ...screw driver in the counter clockwise direction After unlocking the screw take off the battery cover Fig 1 2 Replace the old batteries with new ones and install them on the correct polarity position...

Страница 5: ...uS cm Microsiemens indicates Conductivity measurement 9 Indicates Temperature Coefficient 4 CAL This will be displayed when the unit enters into the calibration mode 10 MAIN DISPLAY For Conductivity...

Страница 6: ...ear key is pressed for 5 seconds all segments of the LCD will be lit up it clears all calibration values stored in the internal memory The unit now requires new conductivity calibration before use Mea...

Страница 7: ...by multiplying conductivity mS cm by a TDS factor The default factor value is 0 65 To change the TDS factor use the Up and Down keys to adjust the value between 0 30 1 00 Press Enter key to save the n...

Страница 8: ...r the values of temperature and conductivity to stabilize for a few seconds Press the Up and Down keys to adjust the reading of the display until it matches the value of the known standard conductivit...

Страница 9: ...e to calibrate the probe of the model 3177 1 Standard solution of 1413 uS cm at 25 Accurately weight out 0 746 grams of research grade dried Potassium Chloride KCL Dissolve in 1000 ml of distilled wat...

Страница 10: ...K constant value input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 0 0 100 0 1 Incorrect standard solution 2 Conductivity cell...

Страница 11: ...S cm 0 1 mg L 1 uS cm 1 mg L 0 01 mS cm 0 01 g L 0 5 F S Conductivity TDS K 10 0 0 1999 uS cm 0 1999 mg L 2 00 19 99 mS cm 2 00 19 99 g L 20 0 199 9 mS cm 20 0 199 9 g L 1 uS cm 1 mg L 0 01 mS cm 0 01...

Страница 12: ...odel and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss...

Отзывы: