2

GENERAL INTRODUCTION

Thank you for selecting the 3177 meter. The 3177 is a precision tool

that measures Conductivity, TDS, Salinity and Temperature. A

built-in microprocessor stores, calculates and compensates for all

parameters related to Conductivity, TDS, Salinity and Temperature

determinations.

This meter has a waterproof IP54 case. The mechanical keys are

highly reliable with tactile and audio feedback. It is powered by six

AAA-size alkaline batteries or with a UL / CE approved AC adapter

(OUTPUT: DC9V). The meter also displays a “BAT” message when

the batteries are in need of replacement. Re-calibration is not

required when power is restored.

The front of the meter has a large LCD that displays Conductivity,

TDS, Salinity and Temperature simultaneously along with user

prompts and mode indicators. The unit prompts the user through

calibration and measurement procedures.

The model 3177 micro-processor allows the user to easily

recalibrate the parameters for the probe. A few keystrokes will

adjust all the parameters for conductivity and will also give the user

the option to select four types (0.01, 0.1, 1.0, 10.0) of probe cell

constant for a better selection of available probes and applications.

And the user can input K value of the cell by keypad directly.

The system simultaneously displays temperature in

along with

either Conductivity, TDS or Salinity. The user can switch back and

forth from all these displays by just pushing a single

“MODE”

key.

Other features include automatic conductivity ranging, automatic

temperature compensation and 50/60 Hz AC noise rejection. This

meter is user-friendly for laboratory application.

INITIAL INSPECTION

Carefully unpack the unit and accessories. Inspect for damages

made in shipment. If any damage is found, notify your

JENCO

representative immediately. All packing materials should be saved

until satisfactory operation is confirmed.

POWER INPUT

The model 3177 can be powered by an UL / CE approved 100 ~ 240

VAC adaptor as well as 6 “AAA” alkaline batteries. Check the label

on the AC adaptor supplied with the instrument to make sure that

the AC line voltage is correct. If the wrong AC adaptor is supplied,

notify your

JENCO

representative immediately.

Содержание VisionPlus 3177

Страница 1: ...0 Operation Manual MODEL 3177 Microcomputer Based Conductivity TDS Salinity Temperature Benchtop Meter 3177...

Страница 2: ...OWER INPUT 2 INSTALLING THE BATTERIES 3 DISPLAY KEYS FUNCTIONS 4 A Display 4 B Keys 5 OPERATIONAL PROCEDURES 6 A Calibration 6 B Conductivity Measurements 8 C Preparing Standard Solutions 8 ERROR DISP...

Страница 3: ...just all the parameters for conductivity and will also give the user the option to select four types 0 01 0 1 1 0 10 0 of probe cell constant for a better selection of available probes and application...

Страница 4: ...screw driver in the counter clockwise direction After unlocking the screw take off the battery cover Fig 1 2 Replace the old batteries with new ones and install them on the correct polarity position...

Страница 5: ...uS cm Microsiemens indicates Conductivity measurement 9 Indicates Temperature Coefficient 4 CAL This will be displayed when the unit enters into the calibration mode 10 MAIN DISPLAY For Conductivity...

Страница 6: ...ear key is pressed for 5 seconds all segments of the LCD will be lit up it clears all calibration values stored in the internal memory The unit now requires new conductivity calibration before use Mea...

Страница 7: ...by multiplying conductivity mS cm by a TDS factor The default factor value is 0 65 To change the TDS factor use the Up and Down keys to adjust the value between 0 30 1 00 Press Enter key to save the n...

Страница 8: ...r the values of temperature and conductivity to stabilize for a few seconds Press the Up and Down keys to adjust the reading of the display until it matches the value of the known standard conductivit...

Страница 9: ...e to calibrate the probe of the model 3177 1 Standard solution of 1413 uS cm at 25 Accurately weight out 0 746 grams of research grade dried Potassium Chloride KCL Dissolve in 1000 ml of distilled wat...

Страница 10: ...K constant value input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 0 0 100 0 1 Incorrect standard solution 2 Conductivity cell...

Страница 11: ...S cm 0 1 mg L 1 uS cm 1 mg L 0 01 mS cm 0 01 g L 0 5 F S Conductivity TDS K 10 0 0 1999 uS cm 0 1999 mg L 2 00 19 99 mS cm 2 00 19 99 g L 20 0 199 9 mS cm 20 0 199 9 g L 1 uS cm 1 mg L 0 01 mS cm 0 01...

Страница 12: ...odel and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss...

Отзывы: