Probe Basic Cell Constant

The main display shows the Cell Constant of the conductivity

probe (calibrated previously or default, the deviation range is

70% ~ 130%). The secondary display shows the current

selected cell constant which is either 0.01, 0.10, 1.00 or 10.0.

Press the

“Up

and

“Down”

keys to select the one you want

to use. Press

“Enter”

key to save the new value and go to the

next calibration parameter.

Standard Solutions Calibration

Immerse the probe in a standard of known conductivity

solution (See section

Preparing Standard Solutions

),

preferably a standard in the middle range of the solutions to

be measured. Immerse the probe (at least 2” to 3” or 5 ~ 7cm

from the tip) without touching the sides of the calibration

container. Shake the probe lightly to remove any air bubbles

trapped in the conductivity cell. The unit will display the

conductivity value. Wait for the values of temperature and

conductivity to stabilize for a few seconds. Press the

Up

and

Down

keys to adjust the reading of the display until it

matches the value of the known standard conductivity solution

at 25

. Press the

Enter

key to save and go to the next

calibration parameter.

K Value Input

(a) The unit will display the conductivity value of the standard

solution with the

CELL K=

staying on.

[Note:

If the

Cell Constant

has been calibrated, please

press the

“Enter”

key to exit calibration and return to

normal operation.

]

(b) Press and hold the

“Up”

or

“Down”

key, the main display

will show the deviation of the conductivity probe. You can

now input the K value (from 70%~130% of the probe basic

cell constant). After releasing the

“Up”

or

“Down”

key,

the unit will display the conductivity value with the

CELL

K=

staying on.

(c) Adjust the K value until the conductivity value displayed on

the LCD matches the value of the known standard

conductivity solution at 25

.

(d) Press

“Enter”

key to save the new

K value

of the cell to

exit calibration and return to normal operation mode.

Содержание VisionPlus 3177

Страница 1: ...0 Operation Manual MODEL 3177 Microcomputer Based Conductivity TDS Salinity Temperature Benchtop Meter 3177...

Страница 2: ...OWER INPUT 2 INSTALLING THE BATTERIES 3 DISPLAY KEYS FUNCTIONS 4 A Display 4 B Keys 5 OPERATIONAL PROCEDURES 6 A Calibration 6 B Conductivity Measurements 8 C Preparing Standard Solutions 8 ERROR DISP...

Страница 3: ...just all the parameters for conductivity and will also give the user the option to select four types 0 01 0 1 1 0 10 0 of probe cell constant for a better selection of available probes and application...

Страница 4: ...screw driver in the counter clockwise direction After unlocking the screw take off the battery cover Fig 1 2 Replace the old batteries with new ones and install them on the correct polarity position...

Страница 5: ...uS cm Microsiemens indicates Conductivity measurement 9 Indicates Temperature Coefficient 4 CAL This will be displayed when the unit enters into the calibration mode 10 MAIN DISPLAY For Conductivity...

Страница 6: ...ear key is pressed for 5 seconds all segments of the LCD will be lit up it clears all calibration values stored in the internal memory The unit now requires new conductivity calibration before use Mea...

Страница 7: ...by multiplying conductivity mS cm by a TDS factor The default factor value is 0 65 To change the TDS factor use the Up and Down keys to adjust the value between 0 30 1 00 Press Enter key to save the n...

Страница 8: ...r the values of temperature and conductivity to stabilize for a few seconds Press the Up and Down keys to adjust the reading of the display until it matches the value of the known standard conductivit...

Страница 9: ...e to calibrate the probe of the model 3177 1 Standard solution of 1413 uS cm at 25 Accurately weight out 0 746 grams of research grade dried Potassium Chloride KCL Dissolve in 1000 ml of distilled wat...

Страница 10: ...K constant value input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 0 0 100 0 1 Incorrect standard solution 2 Conductivity cell...

Страница 11: ...S cm 0 1 mg L 1 uS cm 1 mg L 0 01 mS cm 0 01 g L 0 5 F S Conductivity TDS K 10 0 0 1999 uS cm 0 1999 mg L 2 00 19 99 mS cm 2 00 19 99 g L 20 0 199 9 mS cm 20 0 199 9 g L 1 uS cm 1 mg L 0 01 mS cm 0 01...

Страница 12: ...odel and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss...

Отзывы: