JENCO VisionPlus 3177 Скачать руководство пользователя страница 7

6

OPERATIONAL PROCEDURES

A. Calibration

Calibration setup contains six sections: TDS Constant,

Temperature Coefficient, Temperature Reference, Probe

Basic Cell Constant, Standard Solutions Calibration and K

Value Input. To access these sections:

1.

Connect the conductivity probe to the unit and turn the unit

on.

2.

In conductivity, TDS or salinity mode, allow temperature

reading to stabilize, press

“Cal”

key to enter the calibration

mode. “CAL” icon appears on the LCD.

[Note:

Press

“Enter”

key to accept any values changes in

each section and automatically advance to the next section. If

there are no changes, the unit accepts the current value and

proceeds to the next section.

]

TDS Constant

TDS is determined by multiplying conductivity (mS/cm) by a

TDS factor. The default factor value is 0.65. To change the

TDS factor, use the

“Up”

and

“Down”

keys to adjust the

value between 0.30 ~ 1.00. Press

“Enter”

key to save the

new value and go to the next calibration parameter.

Temperature Coefficient

The unit uses the temperature coefficient to calculate

temperature compensated conductivity. The default value is

1.91%. To change the Temperature Coefficient, use the

“Up”

and

“Down”

keys to adjust the value between 0.00 ~ 4.00%.

Press

“Enter”

key to save the new value and go to the next

calibration parameter.

Temperature Reference

The unit uses the temperature reference value to calculate

temperature compensated conductivity. The default value is

25

. To change the temperature reference, press the

“Up”

and

“Down”

keys to adjust the value between 15 ~ 25

.

Press

“Enter”

key to save the new value and go to the next

calibration parameter.

Содержание VisionPlus 3177

Страница 1: ...0 Operation Manual MODEL 3177 Microcomputer Based Conductivity TDS Salinity Temperature Benchtop Meter 3177...

Страница 2: ...OWER INPUT 2 INSTALLING THE BATTERIES 3 DISPLAY KEYS FUNCTIONS 4 A Display 4 B Keys 5 OPERATIONAL PROCEDURES 6 A Calibration 6 B Conductivity Measurements 8 C Preparing Standard Solutions 8 ERROR DISP...

Страница 3: ...just all the parameters for conductivity and will also give the user the option to select four types 0 01 0 1 1 0 10 0 of probe cell constant for a better selection of available probes and application...

Страница 4: ...screw driver in the counter clockwise direction After unlocking the screw take off the battery cover Fig 1 2 Replace the old batteries with new ones and install them on the correct polarity position...

Страница 5: ...uS cm Microsiemens indicates Conductivity measurement 9 Indicates Temperature Coefficient 4 CAL This will be displayed when the unit enters into the calibration mode 10 MAIN DISPLAY For Conductivity...

Страница 6: ...ear key is pressed for 5 seconds all segments of the LCD will be lit up it clears all calibration values stored in the internal memory The unit now requires new conductivity calibration before use Mea...

Страница 7: ...by multiplying conductivity mS cm by a TDS factor The default factor value is 0 65 To change the TDS factor use the Up and Down keys to adjust the value between 0 30 1 00 Press Enter key to save the n...

Страница 8: ...r the values of temperature and conductivity to stabilize for a few seconds Press the Up and Down keys to adjust the reading of the display until it matches the value of the known standard conductivit...

Страница 9: ...e to calibrate the probe of the model 3177 1 Standard solution of 1413 uS cm at 25 Accurately weight out 0 746 grams of research grade dried Potassium Chloride KCL Dissolve in 1000 ml of distilled wat...

Страница 10: ...K constant value input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 0 0 100 0 1 Incorrect standard solution 2 Conductivity cell...

Страница 11: ...S cm 0 1 mg L 1 uS cm 1 mg L 0 01 mS cm 0 01 g L 0 5 F S Conductivity TDS K 10 0 0 1999 uS cm 0 1999 mg L 2 00 19 99 mS cm 2 00 19 99 g L 20 0 199 9 mS cm 20 0 199 9 g L 1 uS cm 1 mg L 0 01 mS cm 0 01...

Страница 12: ...odel and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss...

Отзывы: