19
C. Relay Action, Relay Setpoint, Hysteresis Mode & Hysteresis Value
If the relay control type is set to HI and the hysteresis mode is
CENTER
, the relay will turn
ON
at [(RELAY SETPOINT) + (0.5 * hysteresis value)], and will turn
OFF
at [(RELAY
SETPOINT) - (0.5 * hysteresis value)].
If the relay control type is set to HI and the hysteresis mode is
EDGE
, the relay will turn
ON
at (RELAY SETPOINT), and will turn OFF at [(RELAY SETPOINT - (hysteresis value)].
If the relay control type is set to LO and the hysteresis mode is
CENTER
, the relay will turn
OFF
at [(RELAY SETPOINT) + (0.5 * hysteresis value)], and will turn
ON
at [(RELAY
SETPOINT) - (0.5 * hysteresis value)].
If the relay type is set to LO and the hysteresis mode is
EDGE
, the relay will turn
OFF
at
[(RELAY SETPOINT) + (hysteresis value)], and will turn
ON
at (RELAY SETPOINT).
H.V.= Hysteresis value (Dead Band)
S.P. = Relay Set point
Relay Action
Hysteresis
mode
Effective RELAY-
ON
Set point
Effective
RELAY-
OFF
Set point
HIGH
CENTER
S.P.+ ½(H.V.)
S.P. -½ (H.V.)
HIGH
EDGE
S.P.
S.P. - H.V.
LOW
CENTER
S.P.-½ (H.V.)
S.P.+½ (H.V.)
LOW
EDGE
S.P
S.P.+ H.V.