RS485-RTU/password settings page
a. Screen illustration
RS485-RTU/PASSWORD
1.PARITY :NONE
SETTINGS
2.RS485-RTU ID:01
3.PASSWORD :0000
1)
PARITY
– This option would let the user choose if the current RS-485 type is
“EVEN”,
“NONE”
or
“ODD”.
2)
RS485-RTU ID
– This is the unique ID/Address for the unit. If the user is connecting
multiple model 6309PDTF or other Jenco models for logging purposes then this
ID/Address must be unique for each connected unit. This ID/Address is the same
address that must be used by the PC program to communicate with this unit.
3)
PASSWORD
- The meter allows the user to set a four digit password. At the
RS485-RTU/password settings page, press
"MODE"
key. The number starts to blink.
Use
"UP"
or
"DOWN"
key to select the desired digit. Press
"ENTER"
key to save
password.
[
Important note:
The password must be entered at password check page to change
settings.
The user is responsible for remembering the password number, otherwise the user
would not be able to calibrate or change all the settings.]
CONTROLLING THE RELAYS
A. Isolation Voltage
The maximum isolation voltage of the relay output contacts is 1500 VDC. The voltage
differential between the relay output contacts and the load should not exceed 1500 VDC.
B. Output Load
The current through the relay output contacts should not exceed 5 Amp at 115 VAC and
2.5 Amp at 230 VAC in order not to cause permanent damage to the relay contacts. This
rating is specified for
resistive
loads only.