Setup
24
6.5 Calibration (Zero Length)
The calibration (test lead compensation) procedure is
performed to set a reference point for cable length to zero
feet/meters. This can be used to compensate for the length
of a test cable – the end of the test cable will be 0
feet/meters after this function is performed. Once calibration
is performed, the ValidatorPRO-NT, Cu retains the
information.
Figure 6.6 Calibration
6.6 Other Settings
The Other Settings button allows the user to be able to turn
on and off the cable frequency response (attenuation) tests
that are run during an auto test. In addition, the user can
turn on and off the auto-save performed after each cable is
tested. Finally, the user can select what format screen shots
are saved in. Using the drop-down dialog box next to the
label ‘Perform Response Tests’, select Yes or No to turn on
or off the cable frequency response test during an ‘Auto
Test’. Using the drop-down dialog box next to the label ‘Turn
On Auto- Save’, select Yes or No to turn on or off the auto-
save feature performed after each cable test during an ‘Auto
Test’. Using the drop-down dialog box next to the label
Содержание ValidatorPRO-NT, Cu
Страница 1: ...User s Guide ValidatorPRO NT Cu www jdsu com know ...
Страница 2: ......
Страница 8: ......
Страница 12: ...About Your ValidatorPRO 8 2 2 Remote Unit Features Figure 2 2 Remote Unit Features ...
Страница 18: ...Integration with Plan Um Software 14 Please refer to the auto test menu section for further details ...
Страница 20: ......
Страница 30: ......
Страница 36: ......
Страница 50: ......
Страница 56: ......
Страница 66: ......
Страница 70: ...www jdsu com know Document Information Doc 21160445 Revision 000 English ...