26
Connector Loss and Backreflection
Ensure that the loss and backreflection levels of all the connectors are low and stable.
Long Cables
Terminate long cables near the DUTs or connectors being tested to eliminate backreflections
from the fiber.
Laser Stability
If the message INPUT HIGH or INPUT LOW is displayed, lower the hinged door of the output
module, connect one of the internal FC/APC connector to the detector, and press the POWER
key. The reading is expected to be both steady and higher than -10 dBm. If the reading is as
expected, but an error message persists, perform the appropriate procedure recommended in
Table 9.
Calibration
If the backreflection or power calibration was not performed properly, set CALBR to 0 dB and
CALP to 0 dB (see the
Calibrating the Meter
section) to restore the factory-set calibration. This
should provide reasonably accurate backreflection and power measurements.
Other
A variety of other possible display messages and symbols are described in Table 9.
Table 9: Other Messages and Symbols
Display Problem
Solution
*
The backreflection or
power measurement is
approaching the range
limit of the RX3 meter.
Follow the setup and measurement
techniques described in the
Preparing
the Meter for Backreflection
Measurements and
Setting Up the
Meter for Loss and Power
Measurements
sections.
<
BR range is very limited.
Follow the setup and measurement
techniques described in the
Preparing
the Meter for Backreflection
Measurements and
Setting Up the
Meter for Loss and Power
Measurements
sections.
TOO MUCH LIGHT
DARK value is too high.
Attach the detector cap to the detector,
and press the DARK key.
INPUT LOW
Laser power is too low.
Check the measurement setup for an
external source that can have entered
the output.
INPUT HIGH
Laser power is too high.
Check the measurement setup for an
external source that can have entered
the output.
(table continued)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание RX3 Series
Страница 5: ...iv Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 12: ...xi Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 13: ...xii Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 51: ...38 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 53: ...40 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...