Maintenance Manual
WBA Series
QA-IF006 R02E Wartungsanleitung WBA_newLogo.doc
Japan Cash Machine Germany GmbH
Page 11 of 16
Quick Adjustment
This procedure is useful for the application on site. It is not as exact as the PC
supported adjustment, but can be carried out easily and without any additional
materials.
Procedure
:
1. turn off the electricity
2. turn on (DIP Switch) No. 5,6,7 and 8 (fig 10)
3. connect the WBA to the power supply
4. insert the black side of the calibration paper (KS-036 ~ KS-039) towards the
middle of the WBA (fig 11)
The acceptor will accept and reject the paper, and will return the paper after the
adjustment.
Fig. 10
Fig.
11