11
2.3.4 Connecting the cables and tubes
Connect the cables and tubes according to the following procedure :
(1) Connecting the cables
1)
Using a digital voltmeter, confirm that the supplied line voltage corresponds to
the voltage shown on the rating plate.
CAUTION
: The line voltage must be confirmed. An outlet can provide an incorrect
voltage due to faulty wiring.
2)
Confirm that the
″
Power
″
switch on the power supply unit is turned OFF.
3)
Plug the cables according to Fig. 2.7.
CAUTION
: Ground the grounding terminal of the power cable.
CAUTION
: For details regarding the wiring and connections of the personal computer
and printer, refer to their respective instruction manuals.
"AC INPUT"
"Amp Unit"
"PS Out"
"PS In"
"Optical Unit"
"Serial"
"CD In"
Detector unit
Amplifier rear panel
Computer
Line voltage
Conecting plate of
monochromator
Fig. 2.7 Connection of cables
(2) Connecting the cooling water tubes (450W light source)
CAUTION
: Bind the tube at the water faucet and the water inlet/outlet ports of the light
source unit, using the supplied tube bands.